China

雷竞技appFluke 120 b series ScopeMeter ® handheld oscillographic table used in industry

  • 雷竞技appFluke 120 b Series Industrial ScopeMeter ® Hand - by Oscilloscopes
  • 120 b Series Industrial ScopeMeter ® Test Tools
  • 雷竞技appFluke 120 b
  • 雷竞技appFluke 120 b

The main features

120 b series ScopeMeter test tools used in industry with innovative features, designed to help technicians to troubleshoot faster and get the required information, so as to ensure the normal operation of the system.Without manual measuring adjustment through the C雷竞技apponnect - and - View ™ trigger and display waveform set technology, and through the Fluke IntellaSet ™ technology automatically displays relevant measuring values.The oscillographic table can be Recorder Event Detect capture and record function is difficult to find the intermittent events, convenient for technicians to view and analysis.

  • Double input digital oscillographic table and multimeter
  • 40 MHz or 20 MHz bandwidth oscillographic table
  • True RMS of two 5000 digital digital multimeter
  • Connect - and - View ™ trigger is easy to use, easy to implement without manual operation
  • IntellaSet ™ technology can intelligently adjust automatically according to the measured signal numerical readout
  • Dual input waveform and meter reading recorder, the long time data can be trend analysis
  • Recorder Event Detect can capture up to 4 kHz repeat elusive intermittent signal waveform
  • Oscillographic strap shielded test leads, resistance and on-off measurement can be performed
  • Resistance, on and off, diode and capacitor meter measurement
  • Power measurement (W, VA, VAR, PF, DPF, Hz)
  • Voltage, current and power harmonic
  • Through BusHealth test check whether industrial network to achieve the physical reference level
  • Save or call data and instrument Settings
  • According to test sequence set, used daily maintenance or commonly used testing process
  • External optical isolation USB interface for transmission, archive and oscillographic analysis data in tables or instrumentation
  • Connect the built-in USB port of the optiona雷竞技appl WiFi adapter, wireless transmission the information to the PC, laptop or Fluke Connect ® * mobile applications
  • Windows ® v雷竞技appersion FlukeView ® ScopeMeter ® software
  • Using robust design, meet the requirements of EN/IEC60529 seismic and 3 g anti-vibration, 30 g IP51 rating protection grade requirements
  • Industrial safety level: the CAT IV 600 V level of security
  • Lithium ion rechargeable batteries, can work continuously for seven hours (charging time is four hours)

* not all models are available in all areas.Please c雷竞技apponsult your local representative Fluke.

The spot
selections10447.00
Tax-inclusive price
Fluke authentic guarantee
With $200 free shipping
Enjoy the return within 7 days

Produc雷竞技appts overview: Fluke 120 b series ScopeMeter ® handheld oscillographic table used in industry

Suggestions to the fluke official flagship store of purchase

Simplify test, electrical and mechanical troubleshooting deeper and more quickly

ScopeMeter ® 120 b series is a compact, rugged, oscillographic table is suitable for industrial electrical and mechanical and electrical equipment troubleshooting and maintenance applications.This test tool set oscillographic table, multimeter and high-speed recorder at a suit, use very convenient.ScopeMeter 120 b series 雷竞技appare also compatible with Fluke Connect ® mobile applications and FlukeView ®, through ScopeMeter software in-depth cooperation key test, data analysis and archived information.

Connect - and - View ™ trigger can be instantaneous and stable display

How difficult oscillographic table users know that triggering.If you use the wrong Settings, waveform capture may lead to instability, and sometimes may lead to wrong measurements.Fluke unique Connect - and - View ™ trigger technique can identify the signal features, and automatically set the trigger correctly, so as to provide a stable, reliable and repeatable.Connect - and - View ™ triggered by the special design, suitable for almost any signal, including electric drive and control signals - without adjusting parameters, don't even need to touch the button.Even for multiple test points to rapid and continuous measurement, can immediately identify the changes and automatically adjust the Settings, so as to provide a stable display.

IntellaSet ™ / automatic reading

Thr雷竞技appough the Fluke IntellaSet ™ technology of automatic reading function using a dedicated algorithm can intelligent analysis of measured waveform, and automatically displayed on the screen of appropriate measurement values, compared with the past, you can more easily obtain the required data.For example, when the measured waveform as the line voltage signal, automatically display Vrms and Hz readings, but if the measured waveforms of square wave, displayed a V and Hz fengfeng value will automatically be reading.Through IntellaSet ™ technology and Connect - and - View ™ trigger automatically, you can not only obtain correct waveform, but also can get the proper reading.The whole process do not touch the button.

Usi雷竞技appng the Fluke Connect - and - View ™ and IntellaSet ™ technology, you can quickly access the required data.

The normal operation of the industrial equipment need reliable power supply, through dual input can get the key power of the measured values.

For single phase or three phase equilibrium system, industrial ScopeMeter ® 120 B series dual input ac + dc RMS voltage that can measure A channel and channel B + dc ac RMS current.Sub雷竞技appsequently, Fluke 125 b will be calculated;Frequency, phase Angle, active power, reactive power (kW) (VA or var), power factor (PF) or displacement power factor (DPF), at the same time, for all phase with the same voltage and current in the three-phase system, also can calculate the power value of the three-phase system.This is can be applied to balance the system load and the resistance.

Easy access to key characteristic to test the power supply power supply.

Harmonic measurement

Harmonic is periodicity of voltage, current, or power sine wave distortion.Power distribution system harmonic is usually caused by nonlinear loads, such as switch dc power supply and the adjustable speed motor drive.Harmonic can lead to transformer, conductor and motor overheating.In harmonic function, the test tool can measure up to 51 times of fundamental wave.At the same time can also to the dc component, THD (total harmonic distortion) and the K factor for measuring, data related to to fully understand the health state of electric load.

Overview of harmonic spectrum with the cursor can be measured relative to the base wave distortion percentage.

Use comprehensive record mode easily diagnose intermittent failures

Hard to find fault is the occasional failure - intermittent events.These intermittent failures may be by improper connection, dust, dirt, corrosion, wire break or damaged joint.Lines and sag or motor startup and shutdown, and other factors can also cause intermittent events, which led to the closing of equipment.When this happens, you may not be able to just found at the scene.But by雷竞技app Fluke ScopeMeter ® test tools can be easily found that these faults.You can plot a minimum and maximum of measurement value of peak, or keeping track waveform.At the same time, through the extensible micro SD memory card can record up to 14 days of operation.The Recorder is also equipped with a Recorder Event Detect, and more powerful, compared with the past, detection and recording of intermittent fault is easier.You only need to set the meter reading or oscillographic table trajectory threshold, the deviation can be marked as a unique event.You don't need to retrieve large amounts of data can easily find out the fault, and can quickly from a tag event to the next, and can also access the complete data set.

Fast display record events, so as to find out and solve the intermittent failures.

Industrial BusHealth Test can determine Industrial bus signal quality

BusHealth Test electrical signal analysis of industrial bus or network, and gives a clear "optimal" on all relevant parameters indication, "weak" or "bad", and displayed next to the actual measured value.Depending on the type of the selected bus (CAN bus, Profi bus, Foundation fieldbus and the RS - 232, etc.), will be compared with standard measurements, or, if there are different requirements for tolerance, CAN set up a unique reference.When the e雷竞技applectrical signal through the network, Fluke 125 b can test the quality of the electrical signals, without having to view the data content.In addition, 125 b, check the signal level and speed, the transition time and distortion, and compare these with the appropriate standards, to help you find errors, such as improper wiring, poor contact, grounding fault or termination is not correct.

Quickly understand the characteristics of the industrial field bus signal of the physical simulation.

A test leads to multiple electric parameters measurement

Only a set of screening test leads to complete the high frequency waveform, instruments and measuring capacitance and resistance, at the same time can also be inspected on off sex.No need to waste time looking for or replace fuses.

In addition to the display waveform, single testing tool can also measure the volt, ohm, amp or capacitance.

Windows ® v雷竞技appersion FlukeView ® ScopeMeter ® software

ScopeMeter 120 b wit雷竞技apph FlukeView ® software, to implement more features:

  • To save equipment color screen shot as a PC
  • The screenshots attached to your report and document
  • Through the ScopeMeter test tool to capture the waveform data, and save it to your PC
  • Create and archive waveform reference, in order to compare easily
  • The waveform data is copied to the spreadsheet to you for a detailed analysis
  • Use the cursor parameters measurement
  • Add user annotations to the instrument setup, and sent to the device and used as instructions for operator reference

Note: the part function is limited to specific ScopeMeter test tool model

• 雷竞技appFluke Connect mobile application compatibility

For automated industrial machinery troubleshooting more difficult than ever before.Only know the test site is far from enough, you also need to know what aspects should be paid attention to - if there is no baseline or unable to consult relevant experts in the field of measurement data, and this is especially difficult.雷竞技appFluke Connect ® Assets wireless software systems and wireless testing tool can provide records of maintenance and easy to read and share accurate data, great help technology to reduce maintenance costs, and increase equipment uptime.Through compare and contrast test point measurement data and trend, you can more fully understand the signal characteristic of a period of time and change.At the same ti雷竞技appme, through the maintenance data is stored in the Fluke Cloud ™, your team members can at any time and place of the required access, so that you can get advice or approval at the scene, make the system more quickly than ever before to restore normal operation.

Product contrast figure 雷竞技appFluke 123 b 雷竞技appFluke 124 b 雷竞技appFluke 125 b
function
A fully functional oscillographic table with two inputs and instrumentation
Oscillographic table bandwidth (MHz) 20 "=" 40 "=" 40 "="
Instrument and oscillographic table recorder
Oscillographic table cursor measurement
The power and harmonic measurement
BusHealth
Including attachments
Now the voltage probe
I400S alternating current clamp

Produc雷竞技appt specifications: Fluke 120 b series of ScopeMeter ® handheld oscillographic table used in industry


Testing wire CatIV, 750 Vrms maximum voltage.A maximum CatIV, 750 Vrms 400 hz
Oscillographic table schema
vertical
Frequency response - dc coupled Not equipped with probe and test lead (with BB120) 123 b: dc maximum 20 MHz (3 dB)
124 b and 125 b: dc biggest 40 MHz (3 dB)
Equipped with 1:1 STL120 to IV shielding test conductor DC - 12.5 MHz (3 dB)/DC - 20 MHz (6 dB)
Equipped with VP41 10:1 probe 123 b: dc maximum 20 MHZ (3 dB)
124 b and 125 b: dc biggest 40 MHz (3 dB)
The frequency response - exchange coupling (if overturned) Not equipped with probe and test lead The < 10 Hz (3 dB)
Equipped with 1:1 STL120 to IV shielding test conductor The < 10 Hz (3 dB)
Equipped with VP41 10:1 probe The < 10 Hz (3 dB)
Rise time, do not contain the probe, 123 b < 17.5 ns
124 b and 125 b < 8.75 ns
The input impedance Not equipped with probe and test lead 1 M Ω / / 20 pF
Equipped with BB120 1 M Ω/m3/24
Equipped with 1:1 STL120 to IV shielding test conductor 1 M Ω / / 230 pF
Equipped with VP41 10:1 probe 5 M Ω / / 15.5 pF
The sensitivity 5 to 200 mV V/div
Analog bandwidth limitations 10 kHz
Display mode A, A, B, B
Maximum input voltage of A and B Direct input, equipped with test leads, or are equipped with VP41 probe 600 Vrms
Equipped with BB120 600 Vrms
Any no-load voltage between terminals and grounding 600 Vrms
The level of
Oscillographic table schema Standard, the single, the rotation
Range (standard) The equivalent sampling 123 b: 20 ns - 500 ns/div,
124 b and 125 b: 10 ns - 500 ns/div
Real-time sampling 1 mu s to 5 s/div
Single (real-time) 1 mu s to 5 s/div
Rotation (real-time) 1 s to 60 s/div
Sampling rate (dual channel sampling at the same time) The equivalent sampling (repeat) Up to 4 GS/s
Real-time sampling 1 mu s to 60 s/div 40 MS/s
The trigger
Update the screen Run automatically, trigger
The data source Type A, B,
A and B sensitivity @ DC - 5 MHz 0.5 div or 5 mV
@ 40 MHz 123 b: 4 div
124 b and 125 b: 1.5 g
@ 60 MHz 123 b: N/A
124 b and 125 b: 4
The slope Positive, negative
Senior oscillographic table function
Display mode normal Capture for 25 ns pulse wave interference, showing similar to continuous analog signal waveform
smooth Inhibition of waveform noise
Shutdown pulse interference Do not capture pulse interference between the sample
envelope Record and display a period of time the waveform of the minimum and maximum values
Automatically set (Connect - and - View ™) Automatic continuous adjusting amplitude, time base, trigger level, trigger delay interval and trigger.Users by adjusting the amplitude, time base or trigger level can be a manual override.

Uncertainty of + 1%.Must can display multiple waveform cycle on the screen.
Double input instrument
When the temperature is 18 ° C to 28 ° C, all the precision of the measurement value in % + last significant figures (reading).
Less than 18 ° C or above 28 ° C, often to reduce or increase 1 ° C and 0.1 x (specific) precision.For voltage measurements, 10:1 probe and a probe
Type A and type B
Dc voltage (VDC)
range 500 mV, 5 V, 50 V, 500 V, 750 V
precision Counting 0.5% plus or minus (+ 5)
Common mode rejection ratio (CMRR) > > 100 dB @ dc, 60 dB @ 50, 60 or 400 Hz
Full scale reading 5000 count
True RMS voltage (V ac and V ac + dc)
range 500 mV, 5 V, 50 V, 500 V, 750 V
Precision range of 5% 100% (dc coupling) DC - 60 Hz ac (V + DC) Plus or minus 1% (+ 10 counts)
1 Hz - 60 Hz ac (V) Plus or minus 1% (+ 10 counts)
Accuracy for the range of 5% 100% (ac or dc coupled) 60 Hz and 20 kHz + + 15 counts (2.5%)
Dc inhibition (VAC) only > 50 dB
Common mode rejection ratio (CMRR) > 100 dB @ dc
> 60 dB @ 50, 60 or 400 Hz
Full scale reading The 5000 count, reading has nothing to do with any signal crest factor.
peak
model Peak, maximum, minimum or fengfeng value
range 500 mV, 5 V, 50 V, 500 V, 2200 V
precision The peak value of maximum peak or minimum accuracy 5% of full scale
The accuracy of fengfeng value 10% of full scale
Full scale reading 500 count
Frequency (Hz)
range 123 b: 1 Hz, 10 Hz and 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz, 10 MHz and 50 MHz
124 b and 125 b: 1 Hz, 10 Hz and 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz, 10 MHz and 70 MHz
Frequency range Continuous automatic setup for 15 Hz (1 Hz) to 50 MHz
The accuracy of 1 Hz to 1 MHz Counting (plus or minus 0.5% + 2)
Full scale reading 10000 times
The RPM
The largest reading 50.00 kRPM
precision Counting (plus or minus 0.5% + 2)
Duty ratio (PULSE)
range Between 2% and 98%
Frequency range Continuous automatic setup for 15 Hz (1 Hz) to 30 MHz
PULSE width (PULSE)
Frequency range Continuous automatic setup for 15 Hz (1 Hz) to 30 MHz
Full scale reading 1000 count
Current (AMP)
Equipped with current clamp range With V ac + dc or ac, dc, V V PEAK is the same
The scale factor 0.1 mV/A, 1 mV/A, 10 mV/A, 100 mV / 400 mV/A, A, V / 1 A, 10 mV/mA
precision With V ac + dc or ac, dc, V V PEAK is the same (forceps combined with current uncertainty)
Temperature (TEMP), with optional temperature probe
range 200 ° C/div (200 ° F/div)
The scale factor 1 mV / ° C and 1 mV / ° F
precision The same as the V dc (plus the temperature probe of uncertainty)
Decibels (dB)
0 dBV 1 V
0 dBm (600 Ω / 50 Ω) 600 Ω or 50 Ω reference conditions for 1 mW
The dB on V + dc ac or dc, V Vac
Full scale reading 1000 count
CREST factor (CREST)
range 1 to 10
Full scale reading 90 count
phase
model A - A - B, B
range 0 to 359 degrees
The resolution of the 1 degree
Power (125 b)
configuration Single-phase/three-phase three-wire balance load (three phase: fundamental component, only limited to "automatic setting" mode)
The power factor (PF) Scope of watt and volt ampere the ratio of 0.00 to 1.00
watt With the input A (v) and type B (amp) is multiplied by the corresponding sampling values to get the RMS values of the readings
Full scale reading 999 count
The VA Vrms x very different
Full scale reading 999 count
Reactive power (VAR) Tick ((VA) squared - W squared)
Full scale reading 999 count
Vpwm
purpose Pulse width modulation (PWM) signal measurement, such as motor drive inverter output signal
The principle of Readings are shown based on the fundamental wave frequency of the number of the whole cycle sampling average voltage effectively
precision Sine wave signals Vrms
Enter A to commonly used
Ohm (Ω)
range 123 b and 124 b 500 Ω Ω, 50 k, 5 k Ω Ω, 500 k Ω, 5 M and 30 M Ω
125 b, Ω 50, 500 Ω Ω, 50 k, 5 k Ω Ω, 500 k Ω, 5 M, 30 M Ω
precision Counting + / - 0.6% (+ 5) to 50 Ω + + 20 count (2%)
Full scale reading 50 Ω count to 5 M Ω - 5000, 30 M Ω - 3000 counts
Measure the current 0.5 mA to 50 nA, decreases with increasing range
The open circuit voltage < 4 V
On and off (Cont)
buzzer < 30 Ω (+ / - 5 Ω), when 50 Ω range
Measure the current 0.5 mA
Short circuit is detected P 1 ms
diode
Measure the voltage 0.5 mA > 2.8 V
When open < 4 V
Measure the current 0.5 mA
polarity The input for A +, COM - pole
Capacitance (CAP)
range NF 50, 500 nF, 5 u F, 50 mu, 500 mu F F
Full scale reading 5000 count
Measure the current 500 nA to 0.5 mA, increases with increasing range
Advanced instrument function
Zero setting According to the reference set the actual value
Automatic stay (type A) Capture and maintain the stability of measurement results.Beeping when measuring stability.Maintain function applied to the meter reading automatically, including AC signal threshold is 1 Vpp, DC signal threshold of 100 mV.
Fixed point Use the attenuation keys enabled


Automatic leveling only)
The cursor readout (124 b and 125 b)
The data source Type A, B,
A single vertical line The average readings, minimum and maximum reading
Since start reading, minimum readings, the average reading of the largest reading and time (in a "rolling" mode, the instrument is in "keep" mode)
The minimum readings, maximum readings since start reading and time (in a "recorder" mode, instrument is in "keep" mode)
The harmonic value of "power quality" mode.
Double vertical line Fengfeng values, time, distance and time distance from bottom of the readings
Average readings, minimum readings, maximum distance and time reading (in a "rolling" mode, the instrument is in "keep" mode)
Double horizontal line High and low and fengfeng is worth reading
Rising or falling time Transition time, 0%, and 100% reading (manual or automatic leveling;Single channel model
precision The same as the precision of oscillographic table
recorder
Recorder captured Meter Recorder Meter reading mode, or in the Scope for consecutive waveform sampling Recorder mode.Information is stored in memory or optional SD card (125 b or 124 b).
Measurement results on the graphic recorder display instrument of maximum and minimum measured values change over time graph, is displayed on the waveform recorder collected samples of the graphics.
Meter reading
Measurement speed Top 2 measurements per second
Record size (minimum, maximum, average) 1 channel 2 M reading
Record the time span 2 weeks
The biggest event number 1024
Waveform recording
Maximum sampling rate Sampling 400 K/SEC
The size of the memory 400 M sampling record time
Memory span 500 mu s/div for 15 minutes
20 ms/div for 11 hours
Record size SD card 1.5 G sample
Record time span SD card 500 mu s/div for 11 hours
20 ms/div for 14 days
The biggest event number 64


(basic 40 Hz - 70 Hz)
Power quality (125 b)
reading W, VA, var, PF, DPF, Hz
W, VA, var range (automatic) 250 W to 250 MW, 625 MW, 1.56 GW
When selected, total harmonic (% r) Counting 2% plus or minus (+ 6)
Selected: when the fundamental wave (% f) Counting 4% plus or minus (+ 4)
DPF 0.00 to 1.00
Power factor 0.00 to 1.00 and 0.04 mm
Frequency range 10.0 Hz to 15.0 kHz
40.0 Hz to 70.0 Hz
The harmonic number DC - 51
Reading/cursor readout Vrms/very different/W Can choose every harmonic of fundamental wave in order to get a single reading
Including base wave frequency, phase Angle and K - factor (in amps or watts)
BusHealth tester 雷竞技app(only Fluke 125 b)
type subtypes agreement
The AS - I NEN - EN50295
CAN ISO - 11898.
Interbus S The RS - 422. EIA - 422.
Modbus The RS - 232. The RS - 232 / EIA - 232
The RS - 485. The RS - 485 / EIA - 485
Foundation Fieldbus H1 61158 class 1, 31.25 kBit
Profibus DP EIA - 485.
PA 61158 class 1

Test tool is closed for 7 hours (screen, waveform and setting) (32 GB)
(10.2 x 5.2 x 2.15 in)雷竞技app FlukeView ® software.
Other specifications
display type 5.7 inch TFT color active matrix
The resolution of the 640 x 480 pixels
Waveform display vertical 10 40 pixels
The level of 12 40 pixels
power external Through BC430 power adapter
The input voltage 10 V DC to 21 V DC
Power consumption 5 W standard
The input connector 5 mm socket
internal Through BP290 battery pack
Battery power 10.8 V lithium ion rechargeable batteries
Working time 7 hours (50%) with brightness
Charging time Testing tool is closed for 4 hours
Allow ambient temperature Charge of 0 to 40 ° C (32 to 104 ° F)
Data is stored Internal storage class store 20 sets of data The micro SD card slot, with optional SD card
mechanical size 259 x 132 x 55 mm
The weight of the 1.4 kg (3.2 lb), including the battery
interface Optical isolation interface Transmission screen capture (bitmap), Settings and data
PC/notebook computer USB interface, can be connected OC4USB photoelectric isolation USB adapters/cables (option), using the Windows ® version
Optional WiFi adapter Fast will screen capture (bitmap), setting and data transmission to the PC/laptop computers, tablets, smart phones, etc.To provide USB port, used to connect to WiFi receiver.For security reasons, do not use USB port through wiring.

Group 1, A class IV, A maximum 750 Vrms 400 Hz
Using the environment
Using the environment MIL - PRF - 28800 - f 2 classes
The temperature Use a battery 0 to 40 ° C (32 to 104 ° F)
Use the power adapter 0 to 50 ° C (32 to 122 ° F)
store - 20 to 60 ° C (4 to 140 ° F)
Working humidity 0 to 10 ° C (32 to 50 ° F) The condensation
10 to 30 ° C (50 to 86 ° F) 95%
30 to 40 ° C (86 to 104 ° F) 75%
40 to 50 ° C (104 to 122 ° F) 45%
store - 20 to 60 ° C (4 to 140 ° F) The condensation
At an altitude of Working altitude is 3 km (10000 feet) The CAT III 600 V
Elevation of 2 km (, 6600 ft) The CAT IV 600 V
store 12 km (40000 feet)
EMC electromagnetic compatibility The international standard IEC 61326-1: industrial, CISPR 11:
South Korea's (KCC) Class A equipment (industrial radio and communication equipment)
The United States (FCC) 47 CFR part 15 B child.According to the provisions of article 15.103 of this product is regarded as tax-free equipment.
Radio with the adapter Frequency range 2412 MHz to 2462 MHz
The output power < 100 mW
Protection grade IP51, consult: EN/IEC60529
Security level general IEC 61010-1: pollution level 2
measurement IEC 61010-2-033: the CAT III 750 V, 600 V/CAT IV
Maximum input voltage of A and B Input directly or through the line input Derating 600 Vrms CAT IV
Equipped with a Banana - to - BNC Adapter BB120 600 Vrms derating
Any no-load voltage between terminals and grounding 600 Vrms Cat

Mode雷竞技appl: Fluke 120 b series ScopeMeter ® handheld oscillographic table used in industry

雷竞技appFLUKE - 125 - b/CN

The Scope Meter ® industrial use handheld oscilloscope (40 MHz)

Include:

  • 雷竞技appFluke 125 b hand-held oscillograph
  • Has the black earth wire armoured test conductor
  • Black test wire (ground)
  • Hook test clamp (red, blue)
  • Banana - to - BNC Adapter (black, 1)
  • Now the voltage probe
  • I 400 s alternating current clamp
  • Angle to the USB adapter
  • Wi Fi * USB adapter
  • Switch mode power supply, adapter/battery charger
  • Rechargeable lithium-ion battery pack
雷竞技appFLUKE - 123 - b/CN

The Scope Meter ® industrial use handheld oscilloscope (20 MHz)

Include:

  • 雷竞技appFluke 123 b hand-held oscillograph
  • Has the black earth wire armoured test conductor
  • Black test wire (ground)
  • Hook test clamp (red, blue)
  • Banana - to - BNC Adapter (black, 1)
  • Angle to the USB adapter
  • Wi Fi * USB adapter
  • Switch mode power supply, adapter/battery charger
  • Rechargeable lithium-ion battery pack
雷竞技appFLUKE - 123 - b/CN/S

The Scope Meter ® industrial use handheld oscilloscope (20 MHz)

Include:

  • 雷竞技appFluke 123 b hand-held oscillograph
  • Has the black earth wire armoured test conductor
  • Black test wire (ground)
  • Hook test clamp (red, blue)
  • Banana - to - BNC Adapter (black, 1)
  • Angle to the USB adapter
  • Wi Fi * USB adapter
  • Switch mode power supply, adapter/battery charger
  • Rechargeable lithium-ion battery pack
  • Portable soft package
  • Magnetic suspension
  • Windows ® v雷竞技appersion Fluke View ® Scope Meter ® software
  • Screen protector
雷竞技appFLUKE - 124 - b/CN

The Scope Meter ® industrial use handheld oscilloscope (40 MHz)

Include:

  • 雷竞技appFluke 124 b hand-held oscillograph
  • Has the black earth wire armoured test conductor
  • Black test wire (ground)
  • Hook test clamp (red, blue)
  • Banana - to - BNC Adapter (black, 1)
  • Now the voltage probe
  • Angle to the USB adapter
  • Wi Fi * USB adapter
  • Switch mode power supply, adapter/battery charger
  • Rechargeable lithium-ion battery pack
雷竞技appFLUKE - 124 - b/CN/S

The Scope Meter ® industrial use handheld oscilloscope (40 MHz)

Include:

  • 雷竞技appFluke 124 b hand-held oscillograph
  • Has the black earth wire armoured test conductor
  • Black test wire (ground)
  • Hook test clamp (red, blue)
  • Banana - to - BNC Adapter (black, 1)
  • Now the voltage probe
  • Angle to the USB adapter
  • Wi Fi * USB adapter
  • Switch mode power supply, adapter/battery charger
  • Rechargeable lithium-ion battery pack
  • Portable soft package
  • Magnetic suspension
  • Windows ® v雷竞技appersion Fluke View ® Scope Meter ® software
  • Screen protector
雷竞技appFLUKE - 125 - b/CN/S

The Scope Meter ® industrial use handheld oscilloscope (40 MHz)

Include:

  • 雷竞技appFluke 125 b hand-held oscillograph
  • Has the black earth wire armoured test conductor
  • Black test wire (ground)
  • Hook test clamp (red, blue)
  • Banana - to - BNC Adapter (black, 1)
  • Now the voltage probe
  • I 400 s alternating current clamp
  • Angle to the USB adapter
  • Wi Fi * USB adapter
  • Switch mode power supply, adapter/battery charger
  • Rechargeable lithium-ion battery pack
  • Portable soft package
  • Magnetic suspension
  • Windows ® v雷竞技appersion Fluke View ® Scope Meter ® software
  • Screen protector