雷竞技app17 b + IND Fluke digital multimeter
The main features
- 600 V CAT III safety design index
- Voltage to upgrade to 6000 words
- Capacitance range increased to 2000 mu F
- The frequency and temperature measurement
- Voltage, resistance, current, capacitance measurement
- Increasing magnetic hanging belt, make the operation more convenient
- Five years warranty
Produc雷竞技appts overview: Fluke 17 b + IND digital multimeter
雷竞技appFluke 17 b + IND
Digital multimeter designed for industrial enterprise users to tailor, in the original precision, durable, safe and easy to use to upgrade again, the stronghold of voltage to upgrade to 6000 words, according to the capacitance range increased to 2000 mu F, upgraded and will test the attachment, increased the magnetic hanging belt make the operation more convenient, in addition to the quality assurance period extended to five years.
Produc雷竞技appt specifications: Fluke 17 b + IND digital multimeter
function | range | The resolution of the | precision | |||||
Ac voltage (40 Hz and 500 Hz)1 |
4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
1.0% + 3 | |||||
Dc voltage | 4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
0.5% + 3 | |||||
Ac voltage (mv) | 400.0 mV | 0.1 mV | 3.0% + 3 | |||||
Dc voltage (mv) | 400.0 mV | 0.1 mV | 1.0% + 10 | |||||
Diode test2 | 2.000 V | 0.001 V | 10% | |||||
resistance (ohm) |
400.0 Ω 4.000 k Ω 40.00 k Ω 400.0 k Ω 4.000 M Ω 40.00 M Ω |
0.1 Ω 0.001 k Ω 0.01 k Ω 0.1 k Ω 0.001 M Ω 0.01 M Ω |
0.5% + 3 0.5% + 2 0.5% + 2 0.5% + 2 0.5% + 2 1.5% + 3 |
|||||
capacitance3 | 40.00 the nF 400.0 the nF 4.000 u F 40.00 u F 400.0 u F 1000 mu F 2000 mu F |
0.01 the nF 0.1 the nF 0.001 u F 0.01 u F 0.1 u F 1 u F 1 u F |
2% + 5 2% + 5 5% + 5 5% + 5 5% + 5 5% + 5 5% + 5 |
|||||
frequency1 Hz (10 Hz to 100 kHz) |
50.00 the H2 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz |
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz |
0.1% + 3 | |||||
Duty ratio1 | Between 1% and 99% | 0.10% | 1%, the typical case4 | |||||
Alternating current mu A (40 Hz and 400 Hz) |
400.0 mu A 4000 mu A |
0.1 mu A 1 mu A |
1.5% + 3 | |||||
Alternating current mA (40 Hz and 400 Hz) |
40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5% + 3 | |||||
Alternating current A (40 Hz and 400 Hz) |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5% + 3 | |||||
Mu A dc current | 400.0 mu A 4000 mu A |
0.1 mu A 1 mu A |
1.5% + 3 | |||||
Dc current mA | 40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5% + 3 | |||||
A dc current | 4.000 A 10.00 A |
0.001 A 0.01 A |
1.5% + 3 | |||||
The temperature | 50 ° C to 400 ° C 0 ° C - 50 ° C - 55 ° C - 0 ° C |
0.1 C | 2% plus or minus 1 ° C Plus or minus 2 ° C 9% plus or minus 2 ° C |
|||||
backlight | - | - | There are | |||||
1All current, frequency and duty ratio range is the range of 1% to 100%.Not less than 1% range of input value specified. 2Usually, the open circuit voltage of 2.0 V, short circuit current < 0.6 mA. 3Specification does not include the test leads capacitance and the capacitance plate caused by error (error range is 40 nF, highest 1.5 nF). 4Usually means the frequency is 50 Hz or 60 Hz, duty ratio between 10% and 90%. |
||||||||
function | Overload protection | Input impedance (nominal value) | Common mode rejection ratio | Normal mode rejection ratio | ||||
Ac voltage | 1000 V1 | > 10 M Ω < 100 pF | In direct current flow, 50 Hz or 60 Hz, greater than 60 dB |
- | ||||
Ac voltage (mv) | 400 mV | > 1 m Ω, < 100 pF | When 50 Hz or 60 Hz, greater than 80 dB | - | ||||
Dc voltage | 1000 V1 | > 10 M Ω < 100 pF | In direct current flow, 50 Hz or 60 Hz, greater than 100 dB | When 50 Hz or 60 Hz, greater than 60 dB | ||||
Dc voltage (mv) | 400 mV | > 1 m Ω, < 100 pF | When 50 Hz or 60 Hz, greater than 80 dB | - | ||||
1106V Hz maximum | ||||||||
General technical indicators | ||||||||
The highest voltage between any terminal and earthing | 1000 V | |||||||
Display (LCD) | 6000 count, update 3 times per second | |||||||
The battery type | 15 a, 2 AA, NEDA IEC LR6 | |||||||
Battery life span | Minimum 500 hours (LED test mode and 50 hours without load cases.Load case, the battery life depends on the kind of tested LED.) | |||||||
The temperature | ||||||||
Working temperature | 0 ° C to 40 ° C | |||||||
Storage temperature | - 30 ° to 60 ° C C | |||||||
Relative humidity | ||||||||
Working humidity | The condensation (< 10 ° C) 10 ° to 30 ° C, C less than or equal to 90% RH. 30 ° C to 40 ° C, less than or equal to 75% |
|||||||
40 M Ω range 10 ° to 30 ° C, C less than or equal to 80% RH. 30 ° C to 40 ° C, less than or equal to 70% |
||||||||
At an altitude of | ||||||||
Working temperature | 2000 m | |||||||
Storage temperature | 12000 m | |||||||
The temperature coefficient of | 0.1 X (specified accuracy) / ° C (18 ° C or > < 28 ° C) | |||||||
The current input fuse protection | 440 mA, 1000 V of fuse,雷竞技app Fluke specified parts only. 11. A, 1000 v of fu雷竞技appse, Fluke specified parts only. |
|||||||
Volume (high x width x length) | 183 x 91 x 49.5 mm | |||||||
The weight of the | 455 g | |||||||
The IP rating | IP 40 | |||||||
security | IEC61010, IEC 61010-1-2-030 CAT III 600 V, the CAT II 1000 V, pollution level 2 | |||||||
Electromagnetic environment | IEC 61326-1: portable | |||||||
Class A equipment (industrial radio and communication equipment)1 1The product meet the requirement of industry (class A) electromagnetic devices, A salesman or A user that should be paid attention to.This equipment is suitable for the business environment, not suitable for family use. |
17 b雷竞技app + models: Fluke IND digital multimeter
雷竞技appFluke 17 b + IND
contains
- 雷竞技app17 b + IND Fluke digital multimeter
- TL75 test pens and, with two protective cap
- Magnetic hanging belt
- 2 AA batteries
- The user manual