雷竞技appFluke 725 multi-function calibrator/verification process
The main features
- Small, streamlined design, easy to carry and handheld
- Solid, reliable and suitable for field use
- Measurement and output shows at the same time, you can also read the input and output value of the instrument
- Measure the volt, ma, platinum resistance (RTD), thermocouple, frequency, and the function of the ohm, can test the sensor/transmitter
- Output/analog voltage, current, thermocouple, platinum resistance (RTD), frequency, resistance and pressure, with calibration transmitter
- Us雷竞技appe the Fluke 700 PXX 29 in any kind of pressure module, can measure/output pressure
- MA output simultaneously measure the pressure for I/P and the valve testing
- Frequency and function of counts per minute (CPM) can support flow meter test
- Automatic step and ramp output linear detection can be performed
- Provide transmitter loop power supply at the same time, the measurement of mA
- Using fast pulse test fast pulse of 10 ms RTD transducer and PLC
- Backlit display, suitable for any lighting condition
- Four AA batteries, battery energy
- Battery door, convenient and replace the battery
Produc雷竞技appts overview: Fluke 725 multi-function calibrator/verification process
Using a simple, powerful!
Pr雷竞技appocess with Fluke 725 multi-function instrument calibrator, you can test and calibration almost all of the process parameters.Ma, it can measure and output voltage, temperature (RTD and thermocouple), frequency, ohm, and pressure signal (external pressure optional modules required).Need to be calibrated transducer?725 split screen, but at the same time observe the input and output.And I/P test for valve, mA in measuring the pressure at the same time can also output.725 with functions of automatic slope and step by step, can be used in remote testing, 25% step function can be used for fast linearity detection.When雷竞技app you receive the Fluke 725 can work immediately.The instrument control rise very simple, no menu, easy to operate.The storage function allows setting more quickly, and very strong.
Produc雷竞技appt specifications: Fluke 725 multi-function calibrator/verification process
Measurement accuracy | ||
Dc voltage | 30.000 V | 0.02% + 2 times count (upper display) |
30.000 V | 0.02% + 2 times count (lower display) | |
100.00 mV | Counting 0.02% + 2 times | |
10.00 mV to 75.00 mV | 0.025% + 1 count (through TC connector) | |
Dc current | 24.000 mA | Counting 0.02% + 2 times |
resistance | 0.0 to 400.0 Ω | 0.15 Ω Ω 0.1 (4), (line 2 and line 3) |
Ω 401 to 1500 | 0.5 Ω line (4), 1 Ω (line 2 and line 3) | |
Ω 1500 to 3200 | 1.5 Ω Ω 1 (4), (line 2 and line 3) | |
frequency | 2.0 to 1000.0 CPM | Count 0.05% + 1 times |
1.0 to 1100.0 Hz | Count 0.05% + 1 times | |
1.00 to 10.00 kHz | Count 0.05% + 1 times | |
The sensitivity | Fengfeng lowest value 1 V | |
pressure | When using any of 29 pressure module accuracy of 0.025% of the range (for detailed specifications, please refer to the options and accessories of pressure module).Provide the following modules: differential pressure, pressure gauge, vacuum degree, absolute pressure, double pressure and high pressure. | |
The output precision | ||
Dc voltage | 100.00 mV | Counting 0.02% + 2 times |
10.000 V | Counting 0.02% + 2 times | |
10.00 mV to 75.00 mV | 0.025% + 1 count (through TC connector) | |
Dc current | 24.000 mA (output) | Counting 0.02% + 2 times |
24.000 mA (simulation) | Counting 0.02% + 2 times | |
resistance | 15.0 to 400.0 Ω | 0.15 Ω (0.15 to 0.5 mA) excitation current, 0.1 Ω (0.5 to 2 mA exciting current) |
Ω 401 to 1500 | 0.5 Ω (exciting current 0.05 to 0.8 mA) | |
Ω 1500 to 3200 | 1 Ω (exciting current 0.05 to 0.4 mA) | |
frequency | 2.0 to 1000.0 CPM | 0.05% |
1.0 to 1100.0 Hz | 0.05% | |
1.00 to 10.00 kHz | 0.25% | |
waveform | 5 V p - p square wave, 0.1 V | |
RTD and thermocouple | ||
Accuracy of measurement | NI - 120. | 0.2 ° c. |
PT - 100 (385). | 0.33 ° c. | |
PT - 100 (393). | 0.3 ° c. | |
PT - 100 (JIS) | 0.3 ° c. | |
PT - 200 (385). | 0.2 ° c. | |
PT - 500 (385). | 0.3 ° c. | |
PT - 1000 (385). | 0.2 ° c. | |
The resolution of the | 0.1 ° c. | |
J | 0.7 ° c. | |
K. | 0.8 ° c. | |
T | 0.8 ° c. | |
E | 0.7 ° c. | |
R | 1.8 ° c. | |
s. | 1.5 ° c. | |
B | 1.4 ° c. | |
L | 0.7 ° c. | |
U | 0.75 ° c. | |
N | 0.9 ° c. | |
The resolution of the | J, K, T, E, L, N, U: 0.1 ° C, 0.1 ° F, R, S, B: 1 ° C, 1 ° F | |
XK | 0.6 ° c. | |
BP | 1.2 ° c. | |
The output precision | NI - 120. | 0.2 ° c. |
PT - 100 (385). | 0.33 ° c. | |
PT - 100 (393). | 0.3 ° c. | |
PT - 100 (JIS) | 0.3 ° c. | |
PT - 200 (385). | 0.2 ° c. | |
PT - 500 (385). | 0.3 ° c. | |
PT - 1000 (385). | 0.2 ° c. | |
The resolution of the | 0.1 ° c. | |
Measurement accuracy for 4 line. | ||
J | 0.7 ° c. | |
K. | 0.8 ° c. | |
T | 0.8 ° c. | |
E | 0.7 ° c. | |
R | 1.4 ° c. | |
s. | 1.5 ° c. | |
B | 1.4 ° c. | |
L | 0.7 ° c. | |
U | 0.75 ° c. | |
N | 0.9 ° c. | |
The resolution of the | J, K, T, E, L, N, U: 0.1 ° C, B, R, S, 1 ° C | |
XK | 0.6 ° c. | |
BP | 1.2 ° c. | |
Technical specifications | ||
Function of the slope | Output function | Voltage, current, resistance, frequency, temperature |
The slope | Slow linear slope, fast linear slope, 25% step slope | |
Circuit power supply function | voltage | 24 V |
precision | 10% | |
Maximum current | 22 mA, short circuit protection | |
Step function | Output function | Voltage, current, resistance, frequency, temperature |
Step by step | The range of 25%, 100% | |
Environmental indicators | ||
Working temperature | - 10 ° C to + 55 ° C | |
Storage temperature | - 20 ° C to 71 ° C | |
Work at an altitude of | 3000 m | |
Safety specifications | ||
approval | EN 61010-93, ANSI/ISA S82.01-1994;1010.1 CAN/CSA C22.2 No: 1992 | |
The mechanical and general specifications | ||
size | 130 x 236 x 61 mm | |
The weight of the | 0.65 kg | |
The battery | 4 AA alkaline batteries | |
Replace the battery | Separate the battery compartment, without breaking the calibration seal operation | |
Port side | Pressure module connector, also used as a remote real-time programming interface |
Mode雷竞技appl: Fluke 725 multi-function calibrator/verification process
Chinese multi-function process calibrator include:
- A set of AC72 alligator clip
- A set of test can fold lines
- A set of TL 75 test line
- The user manual