雷竞技appFluke 753 el multi-function check/verification process signals

  • 雷竞技appFluke 753 Documenting Process Calibrator
  • 雷竞技appFluke 753 Documenting Process Calibrator
  • SS: 753 Documenting Process Calibrator

The main features

More intelligent.More quickly.

New environment photograph harsh environments are more likely to use

753 el verification process signal is a powerful multi-functional recording process verification, the use of the verification, you can download software to create the procedure, lists, and address, or upload data file for printing, and analysis.753 el current signal generator also special built-in powerful HART ® interface, you are now through a separate communication to perform the interface can be used in nearly all daily tasks to complete.

  • Comprehensive upgrade language operating system and concise and friendly operation interface in Chinese, easier for Chinese users to operation, greatly improved the professional on-site calibration and the work efficiency of fault diagnosis.
  • Measure the voltage, mA, RTD, thermocouple, frequency and resistance, can test the sensor, transmitter, and other instruments
  • Output/analog voltage, mA, thermocouple, RTD, frequency, resistance and pressure, can be calibrated transducer
  • During the test using the loop power supply as the transmitter power, at the same time to measure the mA
  • One of t雷竞技apphe module with 29 kinds of Fluke pressure measurement/output pressure
  • Creating and running automated as - found/the as - left the program, or specification document and archive results meet the quality system
  • Download the program and the calibration results for a whole week.
  • Feature-rich, such as automatic step by step, the custom units, during the test user input values, single point and two switch test, square root of DP flow testing, programmable measurement delay, etc
  • Using a simple
  • 3 years warranty
  • Bright Bai Guangshuang display.At the same time read the output parameters and measuring parameters.
  • Multilingual interface
  • Nickel metal hydride rechargeable batteries, using 10 hours continuously.Built-in barometer.
  • Treatment fast pulse type RTD transducer and PLC, shortest pulse is 1 ms.
  • The attached DPC/Track trial software.
  • Compatible with many asset management software package.

Produc雷竞技appts overview: Fluke 753 el multi-function check/verification process signals

Use fluke multi-function verification, accurate finish work!
雷竞技appFluke 753 el multi-function check all the function can meet your need, 743 b and other current signal generator series products will be constantly!

雷竞技appFluke 753 el multi-function checking instrument can be used as a multiple tools - use this tool powerful handheld devices, can output, simulation, and measure the pressure, temperature and electrical signals.Improved new graphical screen, longer service life of lithium ion battery, USB port as well as the package of new attachments will win your love.

Comprehensive upgrade language operating system and concise and friendly operation interface in Chinese, easier for Chinese users to operation, greatly improved the professional on-site calibration and the work efficiency of fault diagnosis.

For log雷竞技appging, Fluke 753 el current signal generator can automatically perform the calibration procedure and capture data.Of course, it also conforms to the ISO 9000, the FDA, EPA and OSHA regulations and so on many strict standards.雷竞技appFluke 753 el recording process verification: more intelligent.More quickly.

View: total pressure calibration solution, all the problem solved

Pressure module

雷竞技appFluke753EL multifunction calibration is almost covers all pressure applications, including gage pressure, differential pressure, double pressure (pressure) composite, absolute pressure and vacuum pressure.
  • In checking instrument Settings there are 10 kinds of different pressure units, with you to specify one of the shows that pressure readings.
  • Solid polyurethane casting shell protection module, can withstand harsh environment and conditions.
  • The inside temperature compensation, ranging from 0 to 50 DHS DHS C C, accuracy is higher.
  • Including the NIST traceable certificate of calibration.Modules can be local calibration, helps to control costs.

A group of 29 optional modules provide pressure calibration and measurement functions.Technical indicators can provide 28 module, basic accuracy of 0.05%.Range from 0-1 inch of water (0-0.25 kPa), until the 0-10000 psi (0-70000 kPa).Other information about the pressure module, please see the pressure module page.

Automated process

For linear transducer, DP flow transmitter and single point and two limit switches, quickly set up powerful automatic calibration procedures.Simply select the corresponding measure and/or output function and fill in the application template.Series 750 el arranged all the rest.It quick test, calculation error and show the final result, highlighting beyond the level of tolerance..

The custom unit

Mapping a one unit to another unit, such as video mapping to the DHS C or DHS F.Allows雷竞技app the use of belt Fluke 80 t - IR temperature probe millivolt output the Fluke 750 el series of accessories, also allows the use of the parts per million or RPM file test of support units.

The value of user input

Technical personnel able to read special equipment such as recorded by the instrument or other output device and/or measurement calibration results.

Limit switch calibration

Program can quickly automatic calibration voltage, current, temperature and pressure of single point and two limit switches.

Differential pressure flow calibration

Programs using the square root function directly instrument calibration DP flow.

Other features
multi-function

Calibration of temperature, pressure, voltage, current, resistance and frequency.Checking instrument which can measure and output, need to use a robust tool, debugging and calibration.

Powerful and easy to use

Easy to learn the menu driven display to guide you to complete any task.A few minutes can begin, rather than a few days.The use of programmable calibration procedures, can create and run automated as - found/the as - left procedures, to ensure that the calibration, consistent, fast.

Records and archives

To support ISO - 90雷竞技app00 standard or specification, Fluke 753 el capture calibration results, the scene without having to carry a pen and paper.Through RS - 232 interface, the result can be transmitted to the PC, so eliminating the manual transcription in the office.

The real hand operation

Very small, convenient in the tool bag and used in narrow space.Rechargeable lithium-ion battery pack enough to support the day.

Solid, reliable,

Depending on the fluke solid design, achieving high accuracy and reliability in harsh environment.Pu molding shell, outer membrane can be used under industrial environment is straightforward.

The bright white light display

All can read the result in any lighting conditions.Three backlit Settings.

Soft key

Can one-click operation task list, automation procedures, scaling, min/Max, step by step and ramp and enhanced functions such as memory.

Three work modes

Measurement, output, or measurement/output at the same time, technical staff just a tool to diagnosis, calibration and maintenance equipment.

Multilingual interface

According to instructions of the languages include: English, French, German, Spanish and Italian.

The built-in calculator

Using the four function with square root, storage, call and executive set instrument or on-site assessment data needed to calculate.Can be used to set the output function for calculating values.Don't need to carry a pen and paper, more do not need to carry other calculators.

A programmable delay measurement

Automation program allows internal calibration instrument slow response.

Produc雷竞技appt specifications: Fluke 753 el multi-function check/verification process signals

Measurement accuracy
Dc voltage Range/resolution 1 year 2 years
100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
Ac voltage (40 Hz - 500 Hz) / 3.000 V to 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
(40 Hz - 500 Hz) / 30.00 V to 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
(40 Hz - 500 Hz) / 300.0 V to 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
Dc current 30.000 mA 0.01% + 5 uA 0.015% + 7 uA
110.00 mA 0.01% + 20 uA 0.015% + 30 uA
resistance 10.000 Ω 0.05% + 50 m Ω 0.07% + 70 m Ω
100.00 Ω 0.05% + 50 m Ω 0.07% + 70 m Ω
1.0000 k Ω 0.05% + 500 m Ω 0.07% + 0.5 Ω
10.000 k Ω 0.1% + 10 Ω 0.15% + 15 Ω
frequency 1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz
1.101/0.001 kHz to 11.000 kHz 0.005 kHz
11.01/0.01 kHz to 50.00 kHz 0.05 kHz
The output precision
1 year 2 years
Dc voltage 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV
1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
Dc current 22.000 mA (output) 0.01% + 0.003 mA 0.02% + 0.003 mA
Current slot (simulation) 0.02% + 0.007 mA 0.04% + 0.007 mA
resistance 10.000 Ω 0.01% + 10 m Ω 0.015% + 15 m Ω
100.00 Ω 0.01% + 20 m Ω 0.015% + 30 m Ω
1.0000 k Ω 0.02% + 0.2 Ω 0.03% + 0.3 Ω
10.000 k Ω 0.02% + 3 Ω 0.03% + 5 Ω
frequency 0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz 0.005 kHz
Technical parameters
Data record function Measurement functions Voltage, current, resistance, frequency, temperature and pressure
Reading speed 1, 2, 5, 10, 20, 30, or 60 times/min
Record the length of the largest 8000 readings (about 30 or 60 times/min, 7980)
Function of the slope Output function Voltage, current, resistance, frequency, temperature
Modulation rate Step 4 per second
Trip to detect Continuity or voltage, output current, cannot be tested for continuity)
Circuit power supply function voltage Optional, 26 V
precision 10%, 22 mA when the minimum to 18 V
Maximum current 25 mA, short circuit protection
Maximum input voltage 50 V dc
Step function Output function Voltage, current, resistance, frequency, temperature
The manual step by step An optional step, use the arrow buttons to change
Automatic step by step Function, startup latency, step time step value, and circulation, etc. Can be completely programming
Environmental indicators
Working temperature - 10 ° C to + 50 ° C
Storage temperature - 20 ° C to + 60 ° C
Dust/waterproof In line with the IP52, IEC 529
Work at an altitude of 3000 metres (9842 feet) above mean sea level
Safety specifications
approval CAN/CSA C22.2 No - 92, ASNI/ISA 1010.1, UL3111 and EN610 S82.01-1994 - and 93
The mechanical and general specifications
size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
The weight of the 1.2 kg (2.7 lb)
The battery Internal lithium ion battery package: 7.2 V, 4400 mAh, 30 Wh
Battery life Usually more than 8 hours
Replace the battery By pressing the touch switch, do not need to open a checking instrument, also need not use other tools
Port side Pressure module connector
The interface connected to the PC USB connector
Optional battery charger/eliminator connection
Data storage capacity The calibration procedure result of 1 week
90 days specifications The 750 series of standard specifications interval is 1 year and 2 years.
A typical 90 - day measure and can through the following method to estimate the output precision: will a year "reading percentage" or "output" percentage divided by 2.
Mesa specifications (use "% of full scale", "words", or "ohm") remains the same.
Temperature, resistance temperature detector
Degrees or reading % - types (alpha) Range ° c. Measurement creates ° C
1 year 2 years
100 Ω Pt (385). - 200-100
Between 100 and 800
0.07 ℃
0.02% + 0.05 ° C
0.14 ℃
0.04% + 0.10 ° C
200 Ω Pt (385). - 200-100
Between 100 and 630
0.07 ℃
0.02% + 0.05 ° C
0.14 ℃
0.04% + 0.10 ° C
500 Ω Pt (385). - 200-100
Between 100 and 630
0.07 ℃
0.02% + 0.05 ° C
0.14 ℃
0.04% + 0.10 ° C
1000 Ω Pt (385). - 200-100
Between 100 and 630
0.07 ℃
0.02% + 0.05 ° C
0.14 ℃
0.04% + 0.10 ° C
100 Ω Pt (3916). - 200-100
Between 100 and 630
0.07 ℃
0.02% + 0.05 ° C
0.14 ℃
0.04% + 0.10 ° C
100 Ω Pt (3926). - 200-100
Between 100 and 630
0.08 ℃
0.02% + 0.06 ° C
0.16 ℃
0.04% + 0.12 ° C
10 Ω Cu (427). - 100-260 0.2 ° c. 0.4 ° c.
120 Ω Ni (672). - 80-260 0.1 ° c. 0.2 ° c.
Source current Output ° c. Allow the current squared
1 year 2 years
1 mA 0.05 ℃
0.0125% + 0.04 ° C
0.10 ℃
0.025% + 0.08 ° C
0.1 mA to 10 mA
500 mu A 0.06 ° c.
0.017% + 0.05 ° C
0.12 ° c.
0.034% + 0.10 ° C
0.1 mA to 1 mA
250 mu A 0.06 ° c.
0.017% + 0.05 ° C
0.12 ° c.
0.034% + 0.10 ° C
0.1 mA to 1 mA
150 mu A 0.06 ° c.
0.017% + 0.05 ° C
0.12 ° c.
0.034% + 0.10 ° C
0.1 mA to 1 mA
1 mA 0.05 ℃
0.0125% + 0.04 ° C
0.10 ℃
0.025% + 0.08 ° C
0.1 mA to 10 mA
1 mA 0.05 ℃
0.0125% + 0.04 ° C
0.10 ℃
0.025% + 0.08 ° C
0.1 mA to 10 mA
3 mA 0.2 ° c. 0.4 ° c. 0.1 mA to 10 mA
1 mA 0.04 ° c. 0.08 ℃ 0.1 mA to 10 mA
1. The RTD measurement for two lines and three line, add 0.4 ° C in the specification.
2. Support the pulse transducer and PLC, short pulse to 1 ms
Temperature, thermocouple
type Output ° c. Measuring ° c. Output ° c.
1 year 2 years 1 year 2 years
E - 250 - - 200 1.3 2.0 0.6 0.9
- 200 - - 100 0.5 0.8 0.3 0.4
- 100-600 0.3 0.4 0.3 0.4
Between 600 and 1000 0.4 0.6 0.2 0.3
N - 200 - - 100 1.0 1.5 0.6 0.9
- 100-900 0.5 0.8 0.5 0.8
Between 900 and 1300 0.6 0.9 0.3 0.4
J - 210 - - 100 0.6 0.9 0.3 0.4
- 100-800 0.3 0.4 0.2 0.3
Between 800 and 1200 0.5 0.8 0.3 0.3
K. - 200 - - 100 0.7 1.0 0.4 0.6
- 100-400 0.3 0.4 0.3 0.4
Between 400 and 1200 0.5 0.8 0.3 0.4
Between 1200 and 1372 0.7 1.0 0.3 0.4
T - 250 - - 200 1.7 2.5 0.9 1.4
- 200 to 0 0.6 0.9 0.4 0.6
0 to 400 0.3 0.4 0.3 0.4
B Between 600 and 800 1.3 2.0 1.0 1.5
Between 800 and 1000 1.0 1.5 0.8 1.2
Between 1000 and 1820 0.9 1.3 0.8 1.2
R - 20 to zero 2.3 2.8 1.2 1.8
0 to 100 1.5 2.2 1.1 1.7
Between 100 and 1767 1.0 1.5 0.9 1.4
s. - 20 to zero 2.3 2.8 1.2 1.8
0 to 200 1.5 2.1 1.1 1.7
Between 200 and 1400 0.9 1.4 0.9 1.4
Between 1400 and 1767 1.1 1.7 1.0 1.5
C 0 to 800 0.6 0.9 0.6 0.9
Between 800 and 1200 0.8 1.2 0.7 1.0
Between 1200 and 1800 1.1 1.6 0.9 1.4
Between 1800 and 2316 2.0 3.0 1.3 2.0
L - 200 - - 100 0.6 0.9 0.3 0.4
- 100-800 0.3 0.4 0.2 0.3
Between 800 and 900 0.5 0.8 0.2 0.3
U - 200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1.0 1.5 0.4 0.6
Between 1000 and 2000 1.6 2.4 0.6 0.9
Between 2000 and 2500 2.0 3.0 0.8 1.2
XK - 200-300 0.2 0.3 0.2 0.5
Between 300 and 800 0.4 0.6 0.3 0.6

Mode雷竞技appl: Fluke 753 el multi-function check/verification process signals

雷竞技appFluke 753 el / 753 plus

Recording process calibration instrument include: document fully functional corrector/process

Current signal generator:

  • BC7240 battery charger
  • BP7240 lithium-ion battery pack
  • DPCTrack 2 ™ software
  • The instructions
  • Calibration can be traced back the NIST report and data
  • Three sets of TP220 test probe and three sets of "long tooth alligator clip
  • Two sets of AC280 clasps clip wire
  • C799 field soft carrying bag
  • The USB communication cables
  • 雷竞技appFluke 753 plus additional include DPCTRACK - REPORTER SOFTWARE