I2000 flex ac current clamp table

  • 雷竞技appFluke i2000 Flex AC Current Clamp

The main features

  • Three range: 20 A, 200 A and 2000 A, 100 mV/A, 10 mV/A and 1 mV/A output
  • Especially suitable for measuring thick wire and inaccessible wires, do not need to disconnect the circuit
  • The maximum measurable wire diameter is 178 mm
  • The CAT III 600 v safety level

Products overview: i2000 flex ac current clamp table

I2000 is a Rogowski principle of ac current probe.With multimeter, recorder or data recorder, can be used to measure current of up to 2000 a.Flexible light measuring head allows the user to quickly and easily for installation in the difficult to close to the area.

Product specifications: i2000 flex ac current clamp table

Electrical specifications
Current range 20 A,
200 A
2000 A
AC RMS
The output sensitivity (exchange coupling) 1 mV
10 mV
100 mV
1 mV/A
The load impedance Minimum 100 k Ω
precision + 25 ° C temperature
The range of + / - 1% (45 Hz - 65 Hz)
linearity From 10% to 100% range
Readings of plus or minus 0.2%
noise 8 mV AC RMS (30 A)
2 mV AC RMS (300 A, 3000 A)
Frequency range 10 Hz to 20 kHz (3 dB)
Phase error "Plus or minus 1 ° (45 Hz - 65 Hz)
Plus or minus 10 ° (20 kHz)
Position the sensitivity From the connection point at the beginning of the cable length > 25 mm
Range of plus or minus 2%
Field effect Starting from the measuring head when cable length > 200 mm
Range of plus or minus 1%
The power supply 2 x AA/MN 1500 / LR6 alkaline batteries, 200 hours of battery, low battery indicator
The temperature coefficient of Readings of plus or minus 0.08% / ° C
Working voltage See section "safety standard"
600 V AC RMS or DC
General technical specifications
Measuring head circumference 610 mm (24) in
Measuring head cable diameter 14.3 mm (0.56 in)
Measuring head bending radius 38.1 mm (1.5 in)
Cable length (measuring head to electronic components) 2 m (6.56 ft)
Output connection 4 mm safety plug molded (integrated)
Working temperature range 0 ° C to + 70 ° C (32 ° F to 158 ° F) (measuring head and electronic components)
Storage temperature range To 40 ° C to + 105 ° C (to 40 ° F to 221 ° F) (measurement)
- 20 ° C to + 85 ° C (4 ° F to 185 ° F) (electronic components)
Working humidity From 15% to 85% (condensate)
The weight of the 180 g (6.34 oz) (measurement)
190 g (6.70 oz) (electronic components)
Safety standards EN 61010-031:2002
EN 61010-2-032:2002
EN 61010-1:2001
600 Vrms, class III, pollution level 2
Using a probe on not insulated wires, shall not exceed 600 V ac RMS or dc, and the frequency is lower than 1 kHz.
EMC standard EN 61326:1998 + A1, A2 and A3
Size (high x width x deep) 116 x 68.5 x 30 mm (4.56 x 2.69 x 1.18 in)

Model: i2000 flex ac current clamp table

雷竞技appFluke i2000 Flex
Ac current clamp table