雷竞技appFluke 1587 c FC insulation multimeter
The main features
- With TrendIt ™ PI/DAR measurement chart
- Usi雷竞技appng a Fluke Connect measurement used for storage
- Usi雷竞技appng a Fluke Connect measurement used for temperature compensation
- Used for accurate measurement of motor driven VFDS low-pass filter
- Insulation test voltage 50 V, 100 V, 250 V, 500 V, 1000 V
- Insulation test: 0.01 M Ω Ω to 2.0 G
- Release the capacitor voltage automatically
- Insulation test smooth reading
- frequency
- capacitance
- Diode test
- The temperature
- Minimum/maximum
- Ac/dc voltage
- Dc voltage (mv)
- Ac/dc current (ma)
- Resistance (0.1 Ω to 50 M Ω)
- continuity
- Remote probe, test lead, alligator clip
- K type thermocouple
- Strong practical hard box
- Automatic shutdown
- Three years warranty
selections6222.00
Produc雷竞技appts overview: Fluke 1587 c FC insulation multimeter
High performance epoxy resin insulating digital multimeter
雷竞技appFluke 1587 c FC insulation multimeter, in a single, compact, portable products, digital insulation tester and function fully, true RMS digital multimeter, in troubleshooting and preventive maintenance as much as possible to play a variety of purposes.
雷竞技appInsulation multimeter by Fluke Fluke 1587 c FC Connect ® measurement applications has added four new powerful diagnostic functions:
- With TrendIt ™ chart of PI/DAR timing than test, insulation can more quickly find that damp and pollution problems.
- By 雷竞技appFluke Connect store the result, but don't have to write a test results, reduce the record error, can save the data, to track history for a long time in the future
- Through the application of temperature compensation helps to establish accurate baseline and related historical data contrast
- To track and trend analysis o雷竞技appf equipment condition, to determine whether insulation function degradation, through the Fluke Connect ® Assets on-site real-time decision-making
Product highlights
- PI/DAR than test regularly
- Live line detection when the voltage is greater than 30 V block insulation test, to enhance the user protection
- Used for accurate measurement of motor driven VFDS low-pass filter
- Automatic release of capacitor voltage, to enhance the user protection
- Insulation test (1587 FC: 0.01 M Ω Ω to 2 G)
- Suitable for many applications of insulation test voltage (1587 FC: 50 V, 100 V, 250 V, 500 V, 1000 V),
- Ac/dc voltage, dc voltage (mv), ac/dc current (ma) and resistance (Ω), on and off
- Capacitance, diode test, temperature and minimum/maximum, frequency (Hz)
- Automatic shutdown to save electricity
- CAT, CAT IV 600 V, 1000 V measurement category III
- Large display with backlight
- Strong practical work hard box allows you to carry all the items you need
- Standard accessories: remote probe, test wiring and probe, alligator clip, K type thermocouple
- Can 雷竞技appchoose Fluke TPAK magnetometer linked hands in convenient operation
- Three standard warranty
Produc雷竞技appt specifications: Fluke 1587 c FC insulation multimeter
5 Hz and 20 kHz kHz5 20 kHz to 100 Hz and 20 kHz kHz5 20 kHz to 100 Hz and 20 kHz kHz5 20 kHz to 100 Hz and 20 kHz kHz5 20 kHz to 100 Hz and 20 kHz 20 kHz to 100 kHz other environment may because of conduction and radiation disturbance and difficult to ensure the electromagnetic compatibility.
The insulation specification | ||
Measured value range | 0.01 M Ω Ω to 2 G | |
The test voltage | 50, 100, 250, 500, 1000 V | |
The test voltage accuracy | + 20% and - 0% | |
Short circuit test current | 1 mA (nominal) | |
Automatic discharge | Discharge time < 0.5 seconds, C = 1 (including F or less | |
Electric circuit testing | Before the initial test, if the terminal voltage > 30 V, inhibition test | |
The biggest capacitive loads | Under 1 (including F load operation | |
Output voltage insulation | ||
50 V (0% to + 20%) | Display range | 0.01 to 6.00 M Ω |
The resolution of the | 0.01 M Ω | |
Test current | 1 mA @ Ω 50 k | |
Resistance to accuracy + + (% readings words) | 3% plus or minus (+ / - 5) | |
Display range | 6.0 to 50.0 M Ω | |
The resolution of the | 0.1 M Ω | |
Test current | 1 mA @ Ω 50 k | |
Resistance to accuracy + + (% readings words) | 3% plus or minus (+ / - 5) | |
100 V (0% to + 20%) | Display range | 0.01 to 6.00 M Ω |
The resolution of the | 0.01 M Ω | |
Test current | 1 mA @ 100 k Ω | |
Resistance to accuracy + + (% readings words) | 3% plus or minus (+ / - 5) | |
Display range | 6.0 to 60.0 M Ω | |
The resolution of the | 0.1 M Ω | |
Test current | 1 mA @ 100 k Ω | |
Resistance to accuracy + + (% readings words) | 3% plus or minus (+ / - 5) | |
Display range | 60 to 100 M Ω | |
The resolution of the | 1 M Ω | |
Test current | 1 mA @ 100 k Ω | |
Resistance to accuracy + + (% readings words) | 3% plus or minus (+ / - 5) | |
250 V (0% to + 20%) | Display range | 0.1 to 60.0 M Ω |
The resolution of the | 0.1 M Ω | |
Test current | 1 mA @ 250 k Ω | |
Resistance to accuracy + + (% readings words) | 1.5% plus or minus (+ / - 5) | |
Display range | 60 to 250 M Ω | |
The resolution of the | 1 M Ω | |
Test current | 1 mA @ 250 k Ω | |
Resistance to accuracy + + (% readings words) | 1.5% plus or minus (+ / - 5) | |
500 V (0% to + 20%) | Display range | 0.1 to 60.0 M Ω |
The resolution of the | 0.1 M Ω | |
Test current | 1 mA @ 500 k Ω | |
Resistance to accuracy + + (% readings words) | 1.5% plus or minus (+ / - 5) | |
Display range | 60 to 500 M Ω | |
The resolution of the | 1 M Ω | |
Test current | 1 mA @ 500 k Ω | |
Resistance to accuracy + + (% readings words) | Plus or minus (1.5% plus or minus five) | |
1000 V (0% to + 20%) | Display range | 0.1 to 60.0 M Ω |
The resolution of the | 0.1 M Ω | |
Test current | 1 mA @ 1 M Ω | |
Resistance to accuracy + + (% readings words) | 1.5% plus or minus (+ / - 5) | |
Display range | 60 to 600 M Ω | |
The resolution of the | 1 M Ω | |
Test current | 1 mA @ 1 M Ω | |
Resistance to accuracy + + (% readings words) | 1.5% plus or minus (+ / - 5) | |
Display range | 0.6 to 2.0 G Ω | |
The resolution of the | 100 M Ω | |
Test current | 1 mA @ 1 M Ω | |
Resistance to accuracy + + (% readings words) | 10% plus or minus (+ / - 3) | |
Electrical technical indicators | ||
Ac voltage measurement | ||
range | ||
600.0 mV | The resolution of the | 0.1 mV |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 5000 Hz + + word reading (%) | Plus or minus 2% (+ 3) | |
6.000 V | The resolution of the | 0.001 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 5000 Hz + + word reading (%) | Plus or minus 2% (+ 3) | |
60.00 V | The resolution of the | 0.01 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 5000 Hz + + word reading (%) | Plus or minus 2% (+ 3) | |
600.0 V | The resolution of the | 0.1 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 5000 Hz + + word reading (%) | Plus or minus 2% (+ 3) creates | |
1000 V | The resolution of the | 1 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 2% (+ 3) | |
The accuracy of 60 Hz - 5000 Hz + + word reading (%) | Plus or minus 2% (+ 3) creates | |
Creates 1 kHz bandwidth | ||
Low pass filter voltage | ||
range | ||
600.0 mV | The resolution of the | 0.1 mV |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 400 Hz + + word reading (%) | + + 3 (2%), - (6% - 3) | |
6.000 V | The resolution of the | 0.001 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 400 Hz + + word reading (%) | + + 3 (2%), - (6% - 3) | |
60.00 V | The resolution of the | 0.01 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 400 Hz + + word reading (%) | + + 3 (2%), - (6% - 3) | |
600.0 V | The resolution of the | 0.1 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 1% (+ 3) | |
The accuracy of 60 Hz - 400 Hz + + word reading (%) | + + 3 (2%), - (6% - 3) | |
1000 V | The resolution of the | 1 V |
Accuracy of 50 Hz or 60 Hz + + word reading (%) | Plus or minus 2% (+ 3) | |
The accuracy of 60 Hz - 400 Hz + + word reading (%) | + + 3 (2%), - (6% - 3) | |
Dc voltage measurement | ||
range | The resolution of the | The accuracy of plus or minus (+ % readings words) |
6.000 V dc | 0.001 V | Plus or minus 0.09% (+ 2) |
60.00 V dc | 0.01 V | Plus or minus 0.09% (+ 2) |
600.0 V dc | 0.1 V | Plus or minus 0.09% (+ 2) |
1000 V dc | 1 V | Plus or minus 0.09% (+ 2) |
The input impedance | 10 M Ω (nominal), < 100 pF | |
Normal mode rejection ratio | > 60 dB @ 50 Hz or 60 Hz | |
Common mode rejection ratio | When dc > 120 dB, 50 Hz or 60 Hz (1 K non-equilibrium) | |
Accuracy is suitable for the + / - 100% range. | ||
Dc millivolt measurement | ||
range | The resolution of the | The accuracy of plus or minus (+ % readings words) |
600.0 mVdc | 0.1 mV | Plus or minus 0.1% (+ 1) |
Dc and ac measurement | ||
Communication 45 Hz to 1000 Hz | ||
range | 400 mA | |
The resolution of the | 0.1 mA | |
The accuracy of plus or minus (+ % readings words) | Plus or minus 1.5% (+ 2) creates | |
The load voltage (typical) | 2 mV/mA | |
range | 60 mA | |
The resolution of the | 0.01 mA | |
The accuracy of plus or minus (+ % readings words) | Plus or minus 1.5% (+ 2) creates | |
The load voltage (typical) | 2 mV/mA | |
dc | ||
range | 400 mA | |
The resolution of the | 0.1 mA | |
The accuracy of plus or minus (+ % readings words) | Plus or minus 0.2% (+ 2) | |
The load voltage (typical) | 2 mV/mA | |
range | 60 mA | |
The resolution of the | 0.01 mA | |
The accuracy of plus or minus (+ % readings words) | Plus or minus 0.2% (+ 2) | |
The load voltage (typical) | 2 mV/mA | |
overload | 600 mA, up to 2 minutes | |
MA input terminal of the fuse protection | 0.44 10 kA mA, 1000 V, IR | |
Ac conversion | The input signal through the exchange coupling and calibration is converted to a sine wave input valid values. | |
Into a true RMS responses, is located in between the range of 5% to 100%.300 mA, the input signal crest factor 3, maximum at 600 mA, linear dropped to 1.5 or less.For non sinusoidal waveforms, maximum crest factor should be combined with 3 + reading (2% + 2% FS) typical values. | ||
Creates 1 kHz bandwidth | ||
Resistance measurement | ||
range | The resolution of the | The accuracy of plus or minus (+ % readings words) creates |
600.0 Ω | 0.1 Ω | Plus or minus 0.9% (+ 2) |
6.000 k Ω | 0.001 k Ω | Plus or minus 0.9% (+ 2) |
60.00 k Ω | 0.01 k Ω | Plus or minus 0.9% (+ 2) |
600.0 k Ω | 0.1 K Ω | Plus or minus 0.9% (+ 2) |
6.000 M Ω | 0.001 M Ω | Plus or minus 0.9% (+ 2) |
50.0 M Ω squared | 0.01 M Ω | Plus or minus 1.5% (+ 3) |
Overload protection | The RMS or dc 1000 V | |
Open circuit voltage | < 8.0 V dc | |
Short circuit current | < 1.1 mA | |
Creates the accuracy is suitable for the range of 0% to 100% Squared up to 80% relative humidity |
||
Diode test | ||
Diode test shows | Display voltage drop: when the nominal test current 1.0 mA for 0.6 V: | |
precision | Plus or minus 2% (+ 3) | |
On and off test | ||
On-off test values | Below 25 Ω resistance test, continuous sound, more than 100 Ω when closed.The biggest reading;1000 Ω | |
The open circuit voltage | < 8.0 V | |
Short circuit current | 1.0 mA typical values | |
Overload protection | True RMS 1000 V | |
The response time | > 1 millisecond | |
Frequency measurement | ||
range | The resolution of the | The accuracy of plus or minus (+ % readings words) |
99.99 Hz | 0.01 Hz | Plus or minus 0.1% (+ 1) |
999.9 Hz | 0.1 Hz | Plus or minus 0.1% (+ 1) |
9.999 kHz | 0.001 kHz | Plus or minus 0.1% (+ 1) |
99.99 kHz | 0.01 kHz | Plus or minus 0.1% (+ 1) |
Frequency counter sensitivity | ||
600.0 mV ac | True RMS ac voltage sensitivity (sine wave) creates | 100.0 mV |
True RMS ac voltage sensitivity (sine wave) creates | 150.0 mV | |
Dc trigger level creates to 20 kHz squared | Do not apply | |
6.0 V | True RMS ac voltage sensitivity (sine wave) creates | 1.0 V |
True RMS ac voltage sensitivity (sine wave) creates | 1.5 V | |
Dc trigger level creates to 20 kHz squared | 400.0 mV and 2.5 V | |
60.0 V | True RMS ac voltage sensitivity (sine wave) creates | 10.0 V |
True RMS ac voltage sensitivity (sine wave) creates | 36.0 V | |
Dc trigger level creates to 20 kHz squared | 1.2 V and 4.0 V | |
600.0 V | True RMS ac voltage sensitivity (sine wave) creates | 100.0 V |
True RMS ac voltage sensitivity (sine wave) creates | - | |
Dc trigger level creates to 20 kHz squared | 12.0 V and 40.0 V | |
1000.0 V | True RMS ac voltage sensitivity (sine wave) creates | 300.0 V |
True RMS ac voltage sensitivity (sine wave) creates | - | |
Dc trigger level creates to 20 kHz squared | 12.0 V and 40.0 V | |
Creates the specified accuracy of maximum input range (up to 1000 V) = 10 times.When low frequency low amplitude noise may affect the accuracy Squared full scale input when applied to 100 kHz |
||
capacitance | ||
range | The resolution of the | The accuracy of plus or minus (+ % readings words) |
1000 nF | 1 nF | Plus or minus 1.2% (+ 2) |
10.00 u F | 0.01 u F | Plus or minus 1.2% (+ 2) |
100.0 u F | 0.1 u F | 90 words plus or minus (+ / - 1.2%) |
9999 mu F | 1 u F | 1.2% plus or minus (+ 90) |
Temperature measurement | ||
range | The resolution of the | The accuracy of plus or minus (+ % readings words) creates |
To 40 ° C to 537 ° C | 0.1 ° c. | Plus or minus (+ / - 1%) |
To 40 ° F to 998 ° F | 0.1 ° F | Plus or minus (+ / - 1% 18) |
Creates the accuracy is suitable for the instrument of environmental temperature change after the measured value of the stable after 90 minutes | ||
General technical indicators | ||
Maximum voltage is commonly applies to any terminal | 1000 V | |
Storage temperature | - 40 ° to 60 ° C - C (to 40 ° F to 140 ° F) | |
Working temperature | - 20 ° C to 55 ° C (4 ° F to 131 ° F) | |
The temperature coefficient of | 28 ° C temperature 18 ° C < or > (< > 64 ° F or 82 ° F), 0.05 x (specified accuracy) / ° C | |
Relative humidity | No condensation | |
10 ° to 30 ° C C (50 ° F to 86 ° F), 0% to 95% | ||
30 ° C to 40 ° C (86 ° F to 104 ° F), 0% to 75% | ||
40 ° C to 55 ° C (104 ° F to 131 ° F), 0% to 40% | ||
vibration | Random, 2 g, 5-500 Hz, based on MIL - PRF - 28800 - f 2 kind of instruments | |
Radio frequency communication | The 2.4 GHz ISM band | |
Radio frequency certification | The FCC: T68 - FBLE, IC: 6627 a - FBLE | |
Electromagnetic compatibility | ||
The international standard | IEC 61326-1: portable electromagnetic environment;IEC 61326-2-2 CISPR group therefore, class A | |
Group 1: equipment to produce and/or use of conduction coupled rf energy, it is necessary for internal operation with the device itself. | ||
Class A: equipment is suitable for the family to use and is not directly connected to the power supply of residential buildings in any equipment of low voltage network. | ||
After the equipment connected to the test object, the launch is likely to exceed the level of CISPR 11.Connect the test lead and/or test probe, the device may not be able to meet the immunity requirements of this standard. | ||
South Korea's (KCC) | Class A equipment (industrial radio and communication equipment) | |
Class A: this product conforms to the requirement of industrial equipment electromagnetic waves, distributor or user should pay attention to this.This equipment is intended for the business environment, rather than the home environment. | ||
The United States (FCC) | 47 CFR part 15 B child.According to the provisions of article 15.103 of this product is regarded as tax-free equipment. | |
Protection grade | IEC 60529: IP40 (aa) | |
security | ||
IEC 61010-1 | Pollution level 2 | |
IEC 61010-2-033 | The CAT III 1000 V, 600 V/CAT IV | |
The battery | 4 AA batteries (IEC LR6) | |
Battery life | Available for instrument use 1000 hours;Insulation test: room temperature, the new alkaline batteries for instrument for at least 1000 times the insulation test.The results of the standard is in 1000 V through 1 M Ω resistance measured, under the condition of the cycle time for 5 seconds, 25 seconds. | |
Size (height x width x length) | 5.0 x 10.0 x 20.3 cm (1.97 x 3.94 x 8.00 in) | |
The weight of the | 550 g (1.2 lb) | |
At an altitude of | Work at an altitude of | 2000 m |
Storage temperature | 12000 m | |
Excess capacity | Except the 110% range, capacitance (100%) | |
Frequency overload protection | < 107 V Hz | |
MA input terminal of the fuse protection | 0.44 A, 1000 V, the IR 10 kA |
Mode雷竞技appl: Fluke 1587 c FC insulation multimeter
FLK - 1587 - FC
雷竞技appFluke 1587 FC insulation multimeter
Include:
- The remote sensor
- The test lead
- Alligator clip
- K type thermocouple
- Hard portable box
- User documentation
Manual + 雷竞技appresources: Fluke 1587 c FC insulation multimeter
- 雷竞技appFluke 1587 | megger insulation test application technology application
- 雷竞技appFluke1587 - Fluke568 ensure the normal operation of variable frequency drive system and motor | application
- Air conditioning compressor motor and electrical insulation resistance testing | application
- Insulation testing can avoid refrigeration equipment accident shutdown | application