With the function雷竞技app of HART communication/diagnosis Fluke 709 h precision circuit verification
The main features
雷竞技appFluke 709 h checking instrument design has a user friendly interface and a Quick - Set rotary encoder knobs.The tool to reduce the measurement of voltage, current, or output and loop power up time.Protective holster is especially suitable for technical staff in hand, large backlit display screen is easy to read, even in the dark, cramped read work area of the screen.709 h verification features:
- 0.01% accuracy of reading
- Compact and durable design, depending on the section 6 standard AAA batteries
- Intuitive user interface, Quick - Set knob can quickly Set up, easy to use
- Built-in HART communication with 250 Ω optional resistor
- 24 V dc circuit of power supply, with mA measurement model (25% ~ 125%)
- Current range resolution of 1, including A voltage range resolution of 1 mV
- All measurement adopts simple double connection
- Automatic shutdown to extend battery life (the longest shutdown time is adjustable to 30 minutes)
- A variable step and ramp time (in seconds)
- Adjustable span selection (0 ~ 20 mA, or 4 ~ 20 mA)
- Tube test (using the % key simulation defined by mA value)
- Fast calibration data.709 h/Track software/cable to. (CSV) or (. TXT) format to record your equipment on all the parameters of the HART transmitter.
- Data can be recorded on a specific transmitter, for easy troubleshooting.Data logger is 1 ~ 60 seconds optional collection interval, can record 9800 records or 99 individual session.Each data sample contains 709 h measurements, all four state of process variables and standard conditions.
In the communication mode, the user can read basic device information, perform diagnostic tests and fine-tuning most HART transmitter calibration.In the past, this operation can only with the help of a dedicated com, a high-end multi-functional checking instrument, or a laptop equipped with HART modems to complete.雷竞技appFluke 709 h can make almost any technician HART device maintenance and support.
selections14468.00
Products overview: with雷竞技app the function of HART communication/diagnosis Fluke 709 h precision circuit verification
雷竞技appFluke 709 h mA loop checking instrument can not only save the time of measurement but also produce high quality results.709 h with HART communication function, and support of the selected a set of general instructions and general application of HART.This is 709 h difference, it is not only an economic, compact circuit verification, and powerful HART communication tool.
Product specifications:雷竞技app with the function of HART communication/diagnosis Fluke 709 h precision circuit verification
Detailed technical indicators | |
function | Output mA, mA, mA, mA reading/circuit and voltage readings. |
range | Current (0 to 24 mA) and voltage (0 to 30 VDC) |
The resolution of the | Current range resolution of 1, including A voltage range resolution of 1 mV |
precision | 0.01% plus or minus 2 digits, all range (@ 23 ° + / - 5 ° C) |
Working temperature range | - 10 ° C to 55 ° C (14 ° F to 131 ° F) |
Humidity range | 10 to 95%, the condensation |
stability | 20 x f.s / ° C (10 ° C - 18 ° C and 28 ° C - 55 ° C) |
display | 128 x 64 pixels, with backlit LCD graphic display, 34 "high digital display |
The power supply | Section 6 AAA alkaline batteries |
Battery life | Usually for 40 hours |
Compliance voltage loop | 24 V direct current 20 ma |
Over voltage protection | 240 V AC |
Overload current protection | 28 mA DC |
Electromagnetic compatibility (EMC) | EN61326 appendix A (portable instrument) |
Size (length x width x height) | 15 x 9 x 3 cm (6 x 3.6 x 1.3 in) |
The weight of the | 0.3 kg (9.5 oz) |
Form a complete set of accessories | NIST traceable certificate of calibration, with data, batteries, test leads, and manual |
Warranty period | Three years, |
With HART communicati雷竞技appon/diagnosis function models: Fluke 709 h precision circuit verification
雷竞技appFluke 709 h
With the function雷竞技app of HART communication/diagnosis Fluke 709 h precision circuit calibration instrument
Contains:
- A 754-8016 alligator clip
- A 75-8014 - x stackable fuses
- TP220-4201 test probe
- Multiple SureGrip hook clamp AC280-5001
- Soft package
- Section six AAA battery (load)
- Cd-rom product manual
- Quick reference guide
- Safety instructions
雷竞技appWith a Fluke 709 h all communication with HART modems, to execute the following command:
- Read the message
- Read the labels, descriptors, calibration date
- Read the PV sensor information
- Read the PV output information
- Read long label
- Write PV range (upper and lower limits)
- Entry/exit fixed current mode
- Zero/migration
- Fine-tuning DAC zero
- Fine-tune the DAC gain