The main features

雷竞技appFluke 725 multi-function process instrument calibrator is powerful and easy to use site calibrator, using the output function measurement or almost any process parameters can test and calibration.
  • Small, streamlined design, easy to carry and handheld
  • Solid, reliable and suitable for field use
  • Measurement and output shows at the same time, you can also read the input and output value of the instrument
  • Measure the volt, ma, platinum resistance (RTD), thermocouple, frequency, and the function of the ohm, can test the sensor/transmitter
  • Output/analog voltage, current, thermocouple, platinum resistance (RTD), frequency, resistance and pressure, with calibration transmitter
  • Us雷竞技appe the Fluke 700 PXX 29 in any kind of pressure module, can measure/output pressure
  • MA output simultaneously measure the pressure for I/P and the valve testing
  • Frequency and function of counts per minute (CPM) can support flow meter test
  • Automatic step and ramp output linear detection can be performed
  • Provide transmitter loop power supply at the same time, the measurement of mA
  • Using fast pulse test fast pulse of 10 ms RTD transducer and PLC
  • Backlit display, suitable for any lighting condition
  • Four AA batteries, battery energy
  • Battery door, convenient and replace the battery

Produc雷竞技appts overview: Fluke 725 multi-function calibrator/verification process

Suggestions to the fluke official flagship store of purchase

Using a simple, powerful!

Pr雷竞技appocess with Fluke 725 multi-function instrument calibrator, you can test and calibration almost all of the process parameters.Ma, it can measure and output voltage, temperature (RTD and thermocouple), frequency, ohm, and pressure signal (external pressure optional modules required).Need to be calibrated transducer?725 split screen, but at the same time observe the input and output.And I/P test for valve, mA in measuring the pressure at the same time can also output.725 with functions of automatic slope and step by step, can be used in remote testing, 25% step function can be used for fast linearity detection.When雷竞技app you receive the Fluke 725 can work immediately.The instrument control rise very simple, no menu, easy to operate.The storage function allows setting more quickly, and very strong.

Produc雷竞技appt specifications: Fluke 725 multi-function calibrator/verification process

Measurement accuracy
Dc voltage 30.000 V 0.02% + 2 times count (upper display)
30.000 V 0.02% + 2 times count (lower display)
100.00 mV Counting 0.02% + 2 times
10.00 mV to 75.00 mV 0.025% + 1 count (through TC connector)
Dc current 24.000 mA Counting 0.02% + 2 times
resistance 0.0 to 400.0 Ω 0.15 Ω Ω 0.1 (4), (line 2 and line 3)
Ω 401 to 1500 0.5 Ω line (4), 1 Ω (line 2 and line 3)
Ω 1500 to 3200 1.5 Ω Ω 1 (4), (line 2 and line 3)
frequency 2.0 to 1000.0 CPM Count 0.05% + 1 times
1.0 to 1100.0 Hz Count 0.05% + 1 times
1.00 to 10.00 kHz Count 0.05% + 1 times
The sensitivity Fengfeng lowest value 1 V
pressure When using any of 29 pressure module accuracy of 0.025% of the range (for detailed specifications, please refer to the options and accessories of pressure module).Provide the following modules: differential pressure, pressure gauge, vacuum degree, absolute pressure, double pressure and high pressure.
The output precision
Dc voltage 100.00 mV Counting 0.02% + 2 times
10.000 V Counting 0.02% + 2 times
10.00 mV to 75.00 mV 0.025% + 1 count (through TC connector)
Dc current 24.000 mA (output) Counting 0.02% + 2 times
24.000 mA (simulation) Counting 0.02% + 2 times
resistance 15.0 to 400.0 Ω 0.15 Ω (0.15 to 0.5 mA) excitation current, 0.1 Ω (0.5 to 2 mA exciting current)
Ω 401 to 1500 0.5 Ω (exciting current 0.05 to 0.8 mA)
Ω 1500 to 3200 1 Ω (exciting current 0.05 to 0.4 mA)
frequency 2.0 to 1000.0 CPM 0.05%
1.0 to 1100.0 Hz 0.05%
1.00 to 10.00 kHz 0.25%
waveform 5 V p - p square wave, 0.1 V
RTD and thermocouple
Accuracy of measurement NI - 120. 0.2 ° c.
PT - 100 (385). 0.33 ° c.
PT - 100 (393). 0.3 ° c.
PT - 100 (JIS) 0.3 ° c.
PT - 200 (385). 0.2 ° c.
PT - 500 (385). 0.3 ° c.
PT - 1000 (385). 0.2 ° c.
The resolution of the 0.1 ° c.
J 0.7 ° c.
K. 0.8 ° c.
T 0.8 ° c.
E 0.7 ° c.
R 1.8 ° c.
s. 1.5 ° c.
B 1.4 ° c.
L 0.7 ° c.
U 0.75 ° c.
N 0.9 ° c.
The resolution of the J, K, T, E, L, N, U: 0.1 ° C, 0.1 ° F, R, S, B: 1 ° C, 1 ° F
XK 0.6 ° c.
BP 1.2 ° c.
The output precision NI - 120. 0.2 ° c.
PT - 100 (385). 0.33 ° c.
PT - 100 (393). 0.3 ° c.
PT - 100 (JIS) 0.3 ° c.
PT - 200 (385). 0.2 ° c.
PT - 500 (385). 0.3 ° c.
PT - 1000 (385). 0.2 ° c.
The resolution of the 0.1 ° c.
Measurement accuracy for 4 line.
J 0.7 ° c.
K. 0.8 ° c.
T 0.8 ° c.
E 0.7 ° c.
R 1.4 ° c.
s. 1.5 ° c.
B 1.4 ° c.
L 0.7 ° c.
U 0.75 ° c.
N 0.9 ° c.
The resolution of the J, K, T, E, L, N, U: 0.1 ° C, B, R, S, 1 ° C
XK 0.6 ° c.
BP 1.2 ° c.
Technical specifications
Function of the slope Output function Voltage, current, resistance, frequency, temperature
The slope Slow linear slope, fast linear slope, 25% step slope
Circuit power supply function voltage 24 V
precision 10%
Maximum current 22 mA, short circuit protection
Step function Output function Voltage, current, resistance, frequency, temperature
Step by step The range of 25%, 100%
Environmental indicators
Working temperature - 10 ° C to + 55 ° C
Storage temperature - 20 ° C to 71 ° C
Work at an altitude of 3000 m
Safety specifications
approval EN 61010-93, ANSI/ISA S82.01-1994;1010.1 CAN/CSA C22.2 No: 1992
The mechanical and general specifications
size 130 x 236 x 61 mm
The weight of the 0.65 kg
The battery 4 AA alkaline batteries
Replace the battery Separate the battery compartment, without breaking the calibration seal operation
Port side Pressure module connector, also used as a remote real-time programming interface

Mode雷竞技appl: Fluke 725 multi-function calibrator/verification process

雷竞技appFluke 725

Chinese multi-function process calibrator include:

  • A set of AC72 alligator clip
  • A set of test can fold lines
  • A set of TL 75 test line
  • The user manual