雷竞技appFluke 725 s multi-function process verification
The main features
- Chinese LCD backlit display, operation is simple
- Measure the volt, ma, platinum resistance (RTD), thermocouple, the function of the frequency, pulse, ohm, can test the sensor/transmitter
- Output/analog voltage, current, thermocouple, platinum resistance (RTD), frequency, pulse, resistance and pressure, with calibration transmitter
- Frequency and pulse counting function can be convenient detection of mass flowmeter, electromagnetic flowmeter metering meter and frequency of pulse output
- High precision pulse input function is used to test maintenance of PLC pulse input card
- Integrated pressure switch testing capabilities, accurately capture the pressure switch to trigger, restore and dead band value
- Us雷竞技appe the Fluke 750 PXX 48 of any kind of pressure module, measurable/output pressure
- High precision measurement and calibration, as high as 0.02%
- Measurement and output shows at the same time, you can also read the input and output value of the instrument
- MA output simultaneously measure the pressure for I/P and the valve testing
- Automatic step and ramp output linear detection can be performed
- Provide transmitter loop power supply at the same time, the measurement of mA
- Save frequently used Settings for later use
- Backlit display, suitable for any lighting condition
- Small, streamlined design, easy to carry and handheld
- Solid, reliable and suitable for field use
- Four AA batteries, battery energy
- Battery door, convenient and replace the battery
Produc雷竞技appts overview: Fluke 725 s multi-function process verification
Using a simple, powerful!
Wi雷竞技appth Fluke 725 s multi-function process verification, you can test and calibration almost all of the process parameters.Ma, it can measure and output voltage, temperature (RTD and thermocouple), frequency, pulse, ohm, and pressure signal (external optional F750P pressure module).Need to repair the calibration transmitter?725 s split screen, but at the same time observe the input and output.And I/P test for valve, mA in measuring the pressure at the same time can also output.725 s with functions of automatic slope and step by step, can be used in remote testing, 25% step function can be used for fast linearity detection.When雷竞技app you receive the Fluke 725 s can work immediately.It is more easy to Chinese display instrument control.The storage function allows setting more quickly, and very strong.Is the maintenance of an ideal maintenance tools
View: total pressure calibration solution, all the problem solved
Produc雷竞技appt specifications: Fluke 725 s multi-function process verification
Accuracy of measurement | ||
Dc voltage | 30.000 V | Upper 0.02% + 2 counts (display) |
30.000 V | 0.02% + 2 counts (lower screen) | |
100.00 mV | 0.02% + 2 counts | |
10.00 mV to 75.00 mV | 0.025% + 1 count (through TC connector) | |
Dc current | 24.000 mA | 0.02% + 2 counts |
resistance | 15.0 to 3200.0 Ω | 0.1 Ω 1.0 Ω |
frequency | 2.0 to 1000.0 CPM | 0.05% + 0.1 count |
1.0 to 1000.0 Hz | 0.05% + 0.1 Hz | |
1.0 to 15.0 kHz | 0.05% + 0.1 kHz | |
10.0 to 15.0 kHz | 0.05% + 1 count | |
The sensitivity | 1 V peak - peak - the minimum | |
pulse | 1 the count | 2.0 CPM to 10000 Hz |
pressure | The accuracy of | Can choose to use any one of 48 pressure module, the module accuracy up to 0.01% |
(for detailed specifications, please refer to the options and accessories of pressure module) | ||
For differential pressure gauge pressure, vacuum pressure, absolute pressure, double pressure and high pressure module. | ||
Accuracy of the output | ||
Dc voltage | 100.00 mV | 0.02% + 2 counts |
10.000 V | 0.02% + 2 counts | |
10.00 mV to 75.00 mV | 0.025% + 1 count (through TC connector) | |
Dc current | 24.000 mA (output) | 0.02% + 2 counts |
24.000 mA (simulation) | 0.02% + 2 counts | |
resistance | range | 15 to 400 Ω |
15 to 400 Ω | ||
Ω 400 to 1500 | ||
Ω 1500 to 4000 | ||
The accuracy of | Reading percentage + base number | |
0.15 Ω | ||
0.1 Ω | ||
0.5 Ω | ||
1.0 Q | ||
Exciting current | 0.15 to 0.5 mA | |
0.5 to 2 ma | ||
0.05 to 0.8 mA | ||
0.05 to 0.4 mA | ||
frequency | 2.0 to 1000.0 CPM | 0.05% |
1.0 to 1100.0 Hz | 0.05% | |
1.00 to 10.00 kHz | 0.05% | |
Waveform: | 5 V p - p square wave, 0.1 V | |
pulse | 1 the count | 2.0 CPM to 10000 Hz |
RTD and thermocouple | ||
Accuracy of measurement and output: | NI - 120. | 0.2 ° c. |
PT - 100 (385). | 0.33 ° c. | |
PT - 100 (392). | 0.30 ° c. | |
PT - 100 (391 - JIS) | 0.30 ° c. | |
PT - 200 (385). | 0.20 ° C (200 ° C to 250 ° C), 0.80 ° C (250 ° C to 630 ° C) | |
PT - 500 (385). | 0.30 ° C (200 ° C to 500 ° C), 0.40 ° C (500 ° C to 630 ° C) | |
PT - 1000 (385). | 0.20 ° c. | |
Resolution: | 0.01 ° c. | |
J: | To 0.7 ° C | |
K: | To 0.8 ° C | |
T: | To 0.8 ° C | |
E: | To 0.7 ° C | |
R: | To 1.4 ° C | |
S: | To 1.5 ° C | |
B: | To 1.4 ° C | |
L: | To 0.7 ° C | |
U: | To 0.75 ° C | |
N: | To 0.9 ° C | |
XK: | To 0.6 ° C | |
BP: | To 1.2 ° C | |
The resolution of the | J, K, T, E, L, N, U: 0.1 ° C, 0.1 ° F B, R, S: 1 ° C, 1 ° F | |
Technical indicators | ||
Oblique wave signal | Output function | Voltage, current, resistance, frequency, pulse and temperature |
slope | Quick slow slope, slope, 25% step slope | |
Circuit power | voltage | 24 V |
The accuracy of | 10% | |
Maximum current | 22 mA, short circuit protection | |
Step signal | Output function | Voltage, current, resistance, frequency, pulse and temperature |
step | The range of 25%, 100% | |
Environmental parameters | ||
Working temperature | - 10 ° C to + 55 ° C | |
Storage temperature | - 20 ° C to + 71 ° C | |
Run the altitude | 3000 m | |
Security technology parameters | ||
Accredited institution | EN 61010-93, took the ANSI/ISA S82.01-1994;1010.1 CAN/CSA C22.2 No: 1992 | |
Machinery and general parameters | ||
size | 47 mm x 200 mm x 96 mm (7.9 in x 3.8 in x 1.9 in) | |
The weight of the | 650 g (23 oz.) | |
The battery | 4 AA alkaline batteries | |
Battery replacement | Split type battery cover, damage to the calibration seal can be loading and unloading | |
Lateral connection | Pressure module connector, can also be used for remote real-time programming |
Mode雷竞技appl: Fluke 725 s multi-function process verification
- A set of AC72 alligator clip
- A set of test can fold lines
- A set of TL 75 test line
- The user manual