China

雷竞技appFluke 754/754 PLUS multi-function process verification - documented fully functional verification - HART

  • 雷竞技appFluke 754 Documenting Process Calibrator - HART
  • 雷竞技appFluke 754 Documenting Process Calibrator - HART
  • 雷竞技appFluke 754 Documenting Process Calibrator - HART
  • SS: 753 Documenting Process Calibrator

The main features

More intelligent.More quickly.

New environment photograph harsh environments are more likely to use

754/754 plus multi-function process verification is a powerful multi-functional recording process calibration instrument, using the instrument, you can download software to create programs, list and instructions, or upload data file for printing, and analysis.754/754 plus special built-in powerful HART ® interface, you are now through a separate communication to perform the interface can be used in nearly all daily tasks to complete.

  • Comprehensive upgrade of language operating system, to provide concise and friendly operation interface in Chinese, easier for Chinese users to operation, greatly improved the professional on-site calibration and the work efficiency of fault diagnosis.
  • Measure the voltage and current (mA), RTD, thermocouple, frequency and resistance, to test the sensor, transmitter, and other instruments.
  • Output/analog voltage and current (mA), thermocouple, RTD, frequency, resistance and pressure calibration transmitter.
  • Thermal signal checking instrument of measuring current at the same time, in the process of testing for transmitter loop power supply.
  • One of the 29 700 PXX pressure module to use fluke/output pressure measurement.
  • Create and run automated calibration before/after the calibration procedure, in order to meet the quality plan or regulations.Records of the results and archived.
  • To download the program and calibration results for up to a week's time.
  • Thermal signals checking instrument has many functions, such as automatic step by step, the custom units, input by the user in the process of test, single point and two point switch, square root of DP flow testing, programmable measurement delay, etc.
  • Easy to use.
  • Three years warranty.
  • High brightness, double screen.
  • At the same time read the output parameters and measuring parameters.
  • Nickel metal hydride rechargeable batteries can use 10 hours continuously.Containing gas pressure gauge.
  • Deal with rapid pulse type RTD transducer and PLC, short pulse to 1 millisecond.
  • The thermal signal check with DPC/Track sampling software.
  • Compatible with a variety of asset management software package.

Produc雷竞技appts overview: Fluke 754/754 PLUS multi-function process verification - documented fully functional verification - HART

雷竞技appFluke 754/754 plus multi-function process verification - documented fully functional verification - HART: more intelligent.More quickly.

View: total pressure calibration solution, all the problem solved

雷竞技appFluke 754/754 plus multi-function process check all the function can meet your need!Combined with the HART communication capabilities to provide integrated calibration instrument of communication.This style is a powerful, reliable performance tools calibration, maintenance and inspection of HART and other instruments of choice.

雷竞技appFluke 754/754 plus multi-function process checking instrument can be used as a multiple tools - using the powerful handheld devices, can output, simulation, and measure the pressure, temperature and electrical signals.Improved new graphical screen, longer service life of lithium ion battery, USB port as well as the package of new accessories will lead you to gain to filmmaking.

雷竞技appFluke 754/754 plus multi-function process verification comprehensive upgrade the operating system, language provides concise and friendly operation interface in Chinese, easier for Chinese users to operation, greatly improved the professional on-site calibration and the work efficiency of fault diagnosis.

For logging, plus 754/754 multi-functional process checking instrument can automatically perform the calibration procedure and capture data.Of course, it also can help you meet ISO 9000, the FDA, EPA and OSHA regulations and so on many strict standards.

HART function
雷竞技appFluke 754/754 plus multi-function process verification is almost complete your now through separate all communicators perform everyday tasks.In fact, it can provide 275 HART communication function of communication.

  • For day-to-day HART, calibration and maintenance, without any external protection shell or other tools.
  • The thermal signal checking instrument provides fast and HART communication.
  • Support common HART transmitter model, compared with any other HART field calibration instrument, has a more specific command support equipment.
  • Can use multiple host, burst mode and branch structure configuration.
  • When to add other instruments and release new version HART, easy to update.
  • Inquiry to determine the device type, manufacturer, model, labels.
  • Reconfigure the dual sensor temperature transmitter of sensor map.
  • Read the HART PV function and intelligent transmitter digital output, mA measurement simulation output at the same time.
  • Read and write HART configuration function, scope of PV point, attenuation and other configuration Settings for site adjustment.
  • Thermal signal verification through reading and writing HART tag field to add intelligent transmitter.

A wide range of HART protocol support

manufacturers Pressure gauge The temperature instrument Coriolis meters
ABB/Kent - Taylor 600 t 658 t1
ABB/Hartmann & Braun Contrans P1,
AS the 800 Series
Endress and Hauser The CERABAR S,
The CERABAR M,
DELTABAR S
TMT 1221TMT 1821,
TMT 1621
Foxboro Eckardt TI/RTT201
Foxboro/Invensys The I/A Pressure
Fuji FCX
FCXAZ
FRC
Honeywell ST3000 STT25T1,
STT25H1
Micro Motion 2000
2000 IS
9701
9712
9739
Moore Products 3441
Rosemount 1151
2088
3001 c
3051, 3051 s
3044 c
644
3144
3244, 3144, p
Siemens SITRANS P DS
SITRANS P ES
SMAR LD301 TT3011
Viatran The I/A Pressure
Wika UNITRANS T32H1
Yokogawa EJA YTA. 110, 310
And 320
1Does not support sensor adjustment

754/754 plus multi-function process verification version 5.7 supports HART protocol contained in the command.With 2 MB of memory, plus 754/754 can support a large number of HART command:

  • General command - provides executive function in all field devices, for example, read the manufacturer and device type, read the main variable (PV) or read the current output percentage and span.
  • Common practice command - provides most (not all) of the field devices commonly used functions, such as read multiple variables, set the attenuation time or execution loop test.
  • Device-specific command - provide specific features unique to the field devices, such as sensors to adjust.

Supports HART operating modes

  • Point to point operation is frequently used mode, this mode will be 754/754 plus is connected to the 4-20 mA circuits of single HART devices.
  • In the multi-branch structure model, the multiple HART instruments can be together.754/754 plus multi-function process verification for each instrument, recognition using the address, and then allows you to select must carry on the calibration of the instruments and related operations.
  • In burst mode, HART instrument to transmit data, without having to wait for the main unit inquiries.During the test or calibration, plus 754/754 transmitter can be taken away from burst mode, and then to restore them to the burst mode.

Pressure module

Almost can be used in any pressure measurement application conditions, including table pressure, differential pressure, double pressure (compound), absolute pressure and vacuum pressure.

  • With you in setting up calibration is specified by any pressure units (up to 10) showed that the pressure readings.
  • Thermal signal check solid polyurethane injection molding shell can protect module, to withstand rough handling and harsh working environment.
  • DHS has 0 to 50 DHS internal C temperature compensation function, to always maintain accurate measurement.
  • Contains the NIST traceable certificate of calibration.
  • Module can be in local calibration, thereby helping to control costs.

29 kinds of optional pressure module series can provide calibration and measurement functions.28 kinds of module basic accuracy specifications of up to 0.05%.Measuring range from 0 to 1 in H20 (0. 25 kPa) to 0-10000 psi (0-70000 kPa).Other information about the pressure modules, please refer to the pressure module page.

The thermal signal checking instrument automation program

Allows for a linear transmitter, DP flow transmitter as well as the single point and two point limit switch quickly set up a powerful automatic calibration procedures.Simply choose the appropriate measurement and/or output function and fill in the application template.The rest are done by 754/754 plus series.It can quickly perform testing, calculation error and display the final result, highlight done almost at the same time.

The custom unit

Allows you to map a unit to another unit, such as the mV is mapped to the DHS C or DHS F.Allo雷竞技appws you to Fluke 754/754 plus multi-function process checking instrument series and the millivolt level output attachments (such as a Fluke 80 t - IR temperature detector), you can also use does not support units, such as the parts per million or revolutions per minute) record test.

The user input values

Enable technicians to record by other equipment (such as equipment instrument panel or only reading) output and/or measurement calibration results.

Limit switch calibration

Procedures for voltage, current, temperature and pressure for the single point and two point limit switch to perform rapid, automatic calibration.

Differential pressure flow meter calibration

The conventional process using the square root function directly to DP flow meter calibration.

Other features
versatile

Calibration of temperature, pressure, voltage, current, resistance and frequency.Because this instrument can measure and can output, so use this powerful tool, you can conduct a comprehensive troubleshooting and calibration.

The thermal signal checking instrument powerful, easy to use

Menu-driven display screen and can be conveniently find throughout the guide when you perform any task.A few minutes to learn, do not need to spend a few days.By programmable calibration program, you can create and run automated calibration before/after the calibration procedure, to ensure the rapid calibration.

The thermal signal verification record results and archived

雷竞技appFluke 754/754 plus multi-function process verification can capture the calibration results, you don't need to use paper and pencil at the scene, to help you perform ISO - 9000 standards or other regulations.By RS - 232 interface, you can to transmit the results to the PC, without having to wait again after back to the shop manual transcription, thereby saving a lot of time.

The thermal signal verification - convenient handheld devices

Very small and can be easily loaded toolkit, also can be used in the cramped space.Using the rechargeable lithium ion battery pack, can use continuously a shift.

Strong and durable, accurate and reliable

雷竞技appFluke design is stable and reliable, even when you are in use in harsh environment with great accuracy and reliability.Cladding forming of urethane can withstand rough handling in industrial environments.

High brightness display

Can read the result under any lighting condition.There are three (3) set back light.

Soft key

Provides one-click access to enhancements, these functions include the task list, automated process, scaling, minimum/maximum climb, step by step, and as well as for memory.

The thermal signal check three operating modes

Measurement, output, or measurement/output at the same time, let the technical personnel using only a tool, can be carried out on the instrument troubleshooting, calibration or maintenance.

Multilingual interface

Display English, French, German, Spanish and Italian.

The built-in calculator

Through four functions, combined with square root, can carry out instrument Settings or storage, recall and field data calculation required for evaluation.Can be used to set the calculated value of the function, of output.With pen and paper, also need not carry extra calculator.

Programmable time delay measurement

Internal automatic program allows for a slower response instrument calibration.

Produc雷竞技appt specifications: Fluke 754/754 PLUS multi-function process verification - documented fully functional verification - HART

Measurement accuracy
Dc voltage Range/resolution 1 year 2 years
100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
Ac voltage (40 Hz - 500 Hz) / 3.000 V to 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
(40 Hz - 500 Hz) / 30.00 V to 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
(40 Hz - 500 Hz) / 300.0 V to 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
Dc current 30.000 mA 0.01% + 5 uA 0.015% + 7 uA
110.00 mA 0.01% + 20 uA 0.015% + 30 uA
resistance 10.000 Ω 0.05% + 50 m Ω 0.07% + 70 m Ω
100.00 Ω 0.05% + 50 m Ω 0.07% + 70 m Ω
1.0000 k Ω 0.05% + 500 m Ω 0.07% + 0.5 Ω
10.000 k Ω 0.1% + 10 Ω 0.15% + 15 Ω
frequency 1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz
1.101/0.001 kHz to 11.000 kHz 0.005 kHz
11.01/0.01 kHz to 50.00 kHz 0.05 kHz
The output precision
1 year 2 years
Dc voltage 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV
1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
Dc current 22.000 mA (output) 0.01% + 0.003 mA 0.02% + 0.003 mA
Current slot (simulation) 0.02% + 0.007 mA 0.04% + 0.007 mA
resistance 10.000 Ω 0.01% + 10 m Ω 0.015% + 15 m Ω
100.00 Ω 0.01% + 20 m Ω 0.015% + 30 m Ω
1.0000 k Ω 0.02% + 0.2 Ω 0.03% + 0.3 Ω
10.000 k Ω 0.02% + 3 Ω 0.03% + 5 Ω
frequency 0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz 0.005 kHz
Technical parameters
Data record function Measurement functions Voltage, current, resistance, frequency, temperature and pressure
Reading speed 1, 2, 5, 10, 20, 30, or 60 times/min
Record the length of the largest 8000 readings (about 30 or 60 times/min, 7980)
Function of the slope Output function Voltage, current, resistance, frequency, temperature
Modulation rate Step 4 per second
Trip to detect Continuity or voltage, output current, cannot be tested for continuity)
Circuit power supply function voltage Optional, 26 V
precision 10%, 22 mA when the minimum to 18 V
Maximum current 25 mA, short circuit protection
Maximum input voltage 50 V dc
Step function Output function Voltage, current, resistance, frequency, temperature
The manual step by step An optional step, use the arrow buttons to change
Automatic step by step Function, startup latency, step time step value, and circulation, etc. Can be completely programming
Environmental indicators
Working temperature - 10 ° C to + 50 ° C
Storage temperature - 20 ° C to + 60 ° C
Dust/waterproof In line with the IP52, IEC 529
Work at an altitude of 3000 metres (9842 feet) above mean sea level
Safety specifications
approval CAN/CSA C22.2 No - 92, ASNI/ISA 1010.1, UL3111 and EN610 S82.01-1994 - and 93
The mechanical and general specifications
size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
The weight of the 1.2 kg (2.7 lb)
The battery Internal lithium ion battery package: 7.2 V, 4400 mAh, 30 Wh
Battery life Usually more than 8 hours
Replace the battery Don't need to open the calibrator for replacement;Don't need to use tools
Port side Pressure module connector
The interface connected to the PC USB connector
Digital instrument (HART) connector
Optional battery charger/eliminator connection
Data storage capacity The calibration procedure result of 1 week
90 days specifications The 750 series of standard specifications interval is 1 year and 2 years.
A typical 90 - day measure and can through the following method to estimate the output precision: will a year "reading percentage" or "output" percentage divided by 2.
Mesa specifications (use "% of full scale", "words", or "ohm") remains the same.
Temperature, resistance temperature detector
Degrees or reading % - types (alpha) Range ° c. Measurement creates ° C
1 year 2 years
100 Ω Pt (385). - 200-100 0.07 ℃ 0.14 ℃
Between 100 and 800 0.02% + 0.05 ° C 0.04% + 0.10 ° C
200 Ω Pt (385). - 200-100 0.07 ℃ 0.14 ℃
Between 100 and 630 0.02% + 0.05 ° C 0.04% + 0.10 ° C
500 Ω Pt (385). - 200-100 0.07 ℃ 0.14 ℃
Between 100 and 630 0.02% + 0.05 ° C 0.04% + 0.10 ° C
1000 Ω Pt (385). - 200-100 0.07 ℃ 0.14 ℃
Between 100 and 630 0.02% + 0.05 ° C 0.04% + 0.10 ° C
100 Ω Pt (3916). - 200-100 0.07 ℃ 0.14 ℃
Between 100 and 630 0.02% + 0.05 ° C 0.04% + 0.10 ° C
100 Ω Pt (3926). - 200-100 0.08 ℃ 0.16 ℃
Between 100 and 630 0.02% + 0.06 ° C 0.04% + 0.12 ° C
10 Ω Cu (427). - 100-260 0.2 ° c. 0.4 ° c.
120 Ω Ni (672). - 80-260 0.1 ° c. 0.2 ° c.
Source current Output ° c. Allow the current squared
1 year 2 years
1 mA 0.05 ℃ 0.10 ℃ 0.1 mA to 10 mA
0.0125% + 0.04 ° C 0.025% + 0.08 ° C
500 mu A 0.06 ° c. 0.12 ° c. 0.1 mA to 1 mA
0.017% + 0.05 ° C 0.034% + 0.10 ° C
250 mu A 0.06 ° c. 0.12 ° c. 0.1 mA to 1 mA
0.017% + 0.05 ° C 0.034% + 0.10 ° C
150 mu A 0.06 C 0.12 C 0.1 mA to 1 mA
0.017% + 0.05 ° C 0.034% + 0.10 ° C
1 mA 0.05 ℃ 0.10 ℃ 0.1 mA to 10 mA
0.0125% + 0.04 ° C 0.025% + 0.08 ° C
1 mA 0.05 ℃ 0.10 ℃ 0.1 mA to 10 mA
0.0125% + 0.04 ° C 0.025% + 0.08 ° C
3 mA 0.2 ° c. 0.4 ° c. 0.1 mA to 10 mA
1 mA 0.04 ° c. 0.08 ℃ 0.1 mA to 10 mA
1. The RTD measurement for two lines and three line, add 0.4 ° C in the specification.
2. Support the pulse transducer and PLC, short pulse to 1 ms
Temperature, thermocouple
type Output ° c. Measuring ° c. Output ° c.
1 year 2 years 1 year 2 years
E - 250 - - 200 1.3 2 0.6 0.9
- 200 - - 100 0.5 0.8 0.3 0.4
- 100-600 0.3 0.4 0.3 0.4
Between 600 and 1000 0.4 0.6 0.2 0.3
N - 200 - - 100 1 1.5 0.6 0.9
- 100-900 0.5 0.8 0.5 0.8
Between 900 and 1300 0.6 0.9 0.3 0.4
J - 210 - - 100 0.6 0.9 0.3 0.4
- 100-800 0.3 0.4 0.2 0.3
Between 800 and 1200 0.5 0.8 0.3 0.3
K. - 200 - - 100 0.7 1 0.4 0.6
- 100-400 0.3 0.4 0.3 0.4
Between 400 and 1200 0.5 0.8 0.3 0.4
Between 1200 and 1372 0.7 1 0.3 0.4
T - 250 - - 200 1.7 2.5 0.9 1.4
- 200 to 0 0.6 0.9 0.4 0.6
0 to 400 0.3 0.4 0.3 0.4
B Between 600 and 800 1.3 2 1 1.5
Between 800 and 1000 1 1.5 0.8 1.2
Between 1000 and 1820 0.9 1.3 0.8 1.2
R - 20 to zero 2.3 2.8 1.2 1.8
0 to 100 1.5 2.2 1.1 1.7
Between 100 and 1767 1 1.5 0.9 1.4
s. - 20 to zero 2.3 2.8 1.2 1.8
0 to 200 1.5 2.1 1.1 1.7
Between 200 and 1400 0.9 1.4 0.9 1.4
Between 1400 and 1767 1.1 1.7 1 1.5
C 0 to 800 0.6 0.9 0.6 0.9
Between 800 and 1200 0.8 1.2 0.7 1
Between 1200 and 1800 1.1 1.6 0.9 1.4
Between 1800 and 2316 2 3 1.3 2
L - 200 - - 100 0.6 0.9 0.3 0.4
- 100-800 0.3 0.4 0.2 0.3
Between 800 and 900 0.5 0.8 0.2 0.3
U - 200 to 0 0.6 0.9 0.4 0.6
0 to 600 0.3 0.4 0.3 0.4
BP 0 to 1000 1 1.5 0.4 0.6
Between 1000 and 2000 1.6 2.4 0.6 0.9
Between 2000 and 2500 2 3 0.8 1.2
XK - 200-300 0.2 0.3 0.2 0.5
Between 300 and 800 0.4 0.6 0.3 0.6

Mode雷竞技appl: Fluke 754/754 PLUS multi-function process verification - documented fully functional verification - HART

雷竞技appFluke 754/754 plus

Multi-function process verification - documented fully functional calibrator - HART

Include:

  • BC7240 battery charger
  • BP7240 lithium batteries battery pack
  • DPCTrack 2 ™ software trial version
  • The instructions
  • Calibration can be traced back the NIST report and data
  • Three sets of TP220 test probe and three sets of "long tooth alligator clip
  • Two sets of AC280 hook clamp
  • C799 field soft tool bag carrying bag
  • The USB communication cables, fluke 754 HHC HART communication cables
  • 雷竞技appFluke 754 plus additional include DPCTRACK - REPORTER SOFTWARE