雷竞技appFluke 753/754 process calibrator firmware upgrade V1.05/1.15 (latest)

雷竞技appFluke 753/754 process calibrator firmware upgrade V1.05/1.15 (latest)

To upgrade to the latest firmware, 753 or 754 operating follow these simple instructions:

  1. Based on your current firmware version of the calibrator, download firmware version 1.05 or 1.15, and save the firmware update file.
    1. 1. If the current firmware version 1.00, 1.01, 1.02, 1.03 or 1.04, pleaseDownload firmware V1.05 (.exe)
    2. If the current firmware version 1.10, 1.11, 1.13 or 1.14, pleaseDownload the V1.15 firmware (.exe)
  2. To ensure that the USB cable to connect the calibrator to a Windows PC, and use the battery charger power supply.Run file and updated according to the instructions.Do not shut down the PC or calibrator during the update.Update is complete, restart the calibrator, check the version number has been updated.The firmware update is completed, do not need to realign the calibrator.
  3. The software licensing of open source software used related optional information available for download.
    1. Open source software license information V1.05. (iso), and is suitable for V1.05 firmware
    2. Open source software license information V1.15. (iso), be applicable to the V1.15 firmware

V1.05 and V1.15:

Add the following tr雷竞技appansducer HART sensor fine-tuning support (Fluke 754) :

  • Rosemount 3051 c, 5.9 and 7.10 Revisions
  • Rosemount 3051 s "C" - Revision 5.7
  • Rosemount 3051 s Diag - 5.3 and 7.4 Revisions
  • Rosemount 2088-5.9 and 7.10 Revisions
  • Rosemount 2090-5.9 and 7.10 Revisions
  • Rosemount 4600 - Revision 5.1

V1.04 and V1.14:

  1. Fixes the fluke dry well control "adjust" before calibration problem
  2. Fixed雷竞技app installation Fluke 750 pa series pressure module of zero

V1.03 and V1.13:

Add the following tr雷竞技appansducer HART sensor fine-tuning support (Fluke 754) :

  • Rosemount 2051-5.9 and 7.10 Revisions
  • Rosemount 3051 s * * * DA2 Advanced Diagnostics - Revision 5.3
  • Rosemount 3144-5.5 and 7.6 Revisions
  • Rosemount 644-5.8 and 7.9 Revisions

If any problems encountered during the update, please start update the application and update again.If the problem still exists, please contact the fluke service mailboxchina-support@雷竞技appfluke.comOr dial the customer service hotline at 400-810-3435.

雷竞技appFluke 753/754 Windows driver

Download the Windows 753/754 start »

The driver supports Windows program to communicate with 753 and 754 through the USB port.Download and run self-extracting installation files.After the installation is complete, 753/754 will be assigned a Windows COM port.If you are using a software package to the serial number, please start Windows device manager, you will see assigned COM port.If you have already installed DPCTrack2, already installed driver, don't have to repeat the installation.For the other 753/754 and communication software, need to install the driver.Some software will automatically find the appropriate COM port, and some require you to manually configure software, need input and 753/754 communication COM ports (in Windows device manager has shown).More information please refer to the specific software.